5 Scope of delivery, transport, set-up and installation
►
20 Hitachi High-Tech Analytical Science
5.5
Establishing optical transparency
The optical components of the instrument are located inside a vacuum chamber in which
a defined negative pressure must be maintained in order to render wavelengths smaller
than 200 nm detectable.
Upon initial installation of the instrument or after prolonged periods of non-use, it is
therefore absolutely necessary to restore this specific vacuum atmosphere to achieve
best measuring results.
Information
Please observe the following:
•
Disable automatic Windows
™
updates (chapter 5.5.1) before starting the pump
cycle to establish optical transparency.
•
Carry out the pump cycle for establishing optical transparency as soon as the
instrument has been switched off or the argon supply has been interrupted.
•
During the pump cycle do not close any of the displayed dialogs of the spectrometer
software and do not switch off the computer.
•
Subsequently perform the UV transparency test (chapter 5.5.3).
5.5.1
Disabling Windows™ updates
In some cases the update function of Windows™ can interrupt the pump cycle. To
prevent this, disable automatic updates as follows:
1.
Open the Windows™ Computer Management.
2.
Select the menu item "Services" (1), then choose "Windows Update" (2) and right-
click your mouse to open the "Properties" (3).
Fig. 5-3
"Computer Management" window
1
2
3
Summary of Contents for OE700 Series
Page 1: ...Operating manual 10022080 Ver 1 0GB 2019 08 14 OE700 series ...
Page 4: ......
Page 8: ...Table of figures VI Hitachi High Tech Analytical Science ...
Page 9: ......
Page 10: ...Table of figures VIII Hitachi High Tech Analytical Science ...
Page 14: ......
Page 20: ......
Page 26: ......
Page 36: ......
Page 46: ......
Page 48: ......
Page 62: ......
Page 66: ......
Page 68: ......
Page 70: ......
Page 72: ......
Page 74: ......
Page 76: ......