background image

-13-

MN1-6403  Rev. 1

4-5. Manipulation of the deflection portion

After inserting the insertion portion into the patient’ s esophagus, flex the tube as indicated

 below in order to observe 

the target region.

Figure 6 

up

down 

right 

left 

UP

Turn the angle knob(for U/D operation) to 

crockwise direction to bend 
the defrectin portion in the in the ultrasound 
irradiation area side.

DOWN
Turn the angle knob(for U/D operation) to 

countercrockwise direction to bend
 the defrectin portion in the in the ultrasound 
irradiation area side.

RIGHT

Turn the angle knob(for L/R operation) to 

crockwise direction to bend the defrectin portion 
to the right side (from acoustic window side 

view).

LEFT

Turn the angle knob(for L/R operation) to 

countercrockwise direction to bend the defrectin 
portion to the left side(from acoustic window side 

view).

DM# 126288 V1

Summary of Contents for MXS2ESLL1 Probe

Page 1: ...e maintenance personnel Please read through this Instruction Manual as well as the separate Instruction Manual Safety MN1 6105 and Cleaning Disinfection and Sterilization MN1 6117 carefully prior to use Keep this Instruction Manual together with the ultrasound diagnostic instrument for any future reference MN1 6403 Rev 1 DM 126288 V1 ...

Page 2: ...ii MN1 6403 Rev 1 DM 126288 V1 ...

Page 3: ...in safety information provided to prevent hazards and injuries to the operator or patient Danger Indicates an imminently hazardous situation which if not avoided will result in death or serious injury to the operator or patient Warning Indicates a potentially hazardous situation which if not avoided could result in death or serious injury to the operator or patient Caution Indicates a potentially ...

Page 4: ...ge operation check 7 3 4 Verification of cleaning disinfection and sterilization 7 3 5 Verification of operation 7 4 Operation 9 4 1 Operation of the bite block 9 4 2 Transducer cover usage precautions 9 4 3 Insertion of Probe 10 4 4 Manipulation of Probe 11 4 5 Manipulation of the deflection portion 13 4 6 Manipulation of the rotation switch 14 4 7 Manipulation of the Lock lever 15 4 8 Temperatur...

Page 5: ...ication based on the degree of protection against electric shock Type BF applied part Classification for protection against ingress of liquids IPX7 Watertight equipment Operation mode Continuous operation Method of sterilization Refer to the separate instruction manual Cleaning Disinfection and Sterilization 1 3 Standard components The standard components of MXS2ESLL1 probe are as follows MXS2ESLL...

Page 6: ... 2 MN1 6403 Rev 1 DM 126288 V1 ...

Page 7: ...ocal Hitachi representative Field of view 90 Scan direction Any direction in 360 relative to insertion direction of insertion portion Frequency 4 0 MHz Cable length 1800 mm Weight 1100g Service life 3 years Applied part See Figure 1 Part treated as Applied part See Figure 1 IPX7 range See Figure 1 Measurement accuracy Refer to the instruction manual of the ultrasound diagnostic instrument External...

Page 8: ...n in the figure below 40 33 50 Unit mm Remarks The tolerance for the dimensions is 10 Figure 2 External view of the bite block 2 3 Transducer cover Use by covering the transducer cover over the insertion portion Transducer cover CIVCO Transducer cover 610 933 Remarks The transducer cover is not included this probe kit DM 126288 V1 ...

Page 9: ...t is listed in Table 2 Figure 3 Name of each parts Insertion portion Proximal head Bite block Connector Deflection portion Rotation Switch A B Cable Angle knob for U D defrection Lock lever Lock knob Probe tip Ultrasonic irradiation part Angle knob for L R defrection Insertion part Rotation Switch C Ring DM 126288 V1 ...

Page 10: ...tch C pushing this switches makes rotation of the diagnostic image Cable Cable transfers electric input output signals Connector The connector is the part which is connected to the ultrasound diagnostic instrument Ring This part is used for hanging the proximal head Bite block This is put in patient s mouth so that keep the path of the probe inserted into the body Caution Do not pull bend twist or...

Page 11: ...e the lock lever and lock knob to make sure there are no abnormalities in the curvature holding or releasing functions 3 3 Diagnostic image operation check Push the switches in each direction of the diagnostic image as far as it will go on the displayed screens and confirm the following Make sure the diagnostic image rotates smoothly in all directions 3 4 Verification of cleaning disinfection and ...

Page 12: ... 8 MN1 6403 Rev 1 DM 126288 V1 ...

Page 13: ...ly or gel this may cause exterior deterioration Use Hitachi approved transducer cover only If use transducer cover which not reccomended can cause an adverse reaction by the body of the patient Check that the transducer cover is sterilized Use of an infected item could spread infection to the patient Inside of transducer cover use the sterilized echo jelly gel which is attached transducer cover If...

Page 14: ...n accordance with the supplied documentation can result in a serious or moderate injury equipment breakdown or physical damage that impairs operation Do not use on the eyes This probe is not intended for use on the eyes The acoustic output can have an adverse effect on the eyes Do not attempt to disassemble modify or repair the equipment Electric shock or other unforeseen accidents could result Co...

Page 15: ...ather preferably under x ray monitoring to check the state of the probe carefully push the probe until its tip comes into the stomach and straighten the bend by operating the angle knob before pulling back the probe Forcing it to move can injure the patient Occuring J turn The probe gets bent in the esophageal Push the probe until stomach and straighten Keep the angle knob in the free position dur...

Page 16: ...it can reduce the performance or cause smoke or fire If anything unusual occurs immediately stop using it and contact one of our offices and or distributor s offices listed on the back cover Use move and transport the equipment under the environmental conditions specified in this manual Otherwise it may be damaged Do not use with a defibrillator This should result in the probe damaged If you have ...

Page 17: ... portion in the in the ultrasound irradiation area side DOWN Turn the angle knob for U D operation to countercrockwise direction to bend the defrectin portion in the in the ultrasound irradiation area side RIGHT Turn the angle knob for L R operation to crockwise direction to bend the defrectin portion to the right side from acoustic window side view LEFT Turn the angle knob for L R operation to co...

Page 18: ... the diagnostic image using the method shown in the figure below in order to observe the region of interest The angle indication is a reference level B Mode Switch A Turn the image to direction Switch B Turn the image to direction Switch C Turn the image by step to direction Switch A Switch B Switch C Direction Direction Reference image DM 126288 V1 ...

Page 19: ...on into the patient s esophagus flex the portion as indicated below in order to observe the target region Lock of angle knob for U D operation Turn the lock lever to fix an angle of the diflection portion and opposite direction to free Lock of angle knob for L R operation Turn the lock knob to fix an angle of the diflection portion and opposite direction to free direction direction Reference image...

Page 20: ...he patient s esophagus Be sure to immediately wash and properly cleaning disinfection after use Warning Keep the angle knob in the free position during an operation of pulling out Removing the deflection portion while locked may injure the patient When blood attaches to Probe or accessories please do cleaning and sterilization If reuse without sterilizing there might be the infection to the patien...

Page 21: ... 17 MN1 6403 Rev 1 DM 126288 V1 ...

Page 22: ... 18 MN1 6403 Rev 1 This page intentionally left blank DM 126288 V1 ...

Page 23: ... 19 MN1 6403 Rev 1 This page intentionally left blank DM 126288 V1 ...

Page 24: ...Manufacturer Hitachi Ltd 2 16 1 Higashi Ueno Taito ku Tokyo 110 0015 Japan Contact 81 3 6284 3668 http www hitachi com businesses healthcare index html Overseas Offices Hitachi Medical Systems GmbH Otto von Guericke Ring 3 D 65205 Wiesbaden Germany EU Importer Hitachi Medical Systems Europe Holding AG Address Sumpfstrasse 13 CH 6300 Zug Switzerland DM 126288 V1 ...

Reviews: