8. Press 

EVAC 

on 

SC AIRLOCK VALVE DISPLAY 

- this will pump the specimen 

exchange chamber. 
9.  Observe SEC vacuum display as status lights show vacuum goes from AIR to LOW to 
HIGH. 
10.  When SEC vacuum is FilCH (green light is on), open airlock door, by turning airlock level 
anticlockwise about one half turn. 

DO NOT FORCE THIS LEVER, 

it requires minimal 

effort to open. 
11.  When airlock door is open, fully insert specimen exchange rod into specimen chamber. 
Watch the specimen being inserted and ensure that it sits properly inside the specimen chamber. 
When rod is fully inserted, unscrew rod from sample by turning rod anticlockwise. 
12.  When rod is unscrewed, withdraw the rod into the specimen exchange chamber. 

13.   Close airlock door by turning airlock level clockwise about one half turn. 
 

C  

TURNING THE BEAM ON – 

GETTING AN IMAGE

 

1.  Select the required accelerating voltage Press 

PF1 

Vacc 

2.  Using numeric keyboard key in desired accelerating voltage e.g. 

5kv. 

3. 

Press 

ENTER 

4.  Microscope will automatically go to 5kV. 

5.

 

Press 

PF16 EXIT 

to clear screen  

 

6.

 

Press the ON bunon on the HV display (middle button). The red left hand button will flash 

while the high tension comes on. When the beam current has been reached the light will stop 

flashing. 

7.  Set magnification to  IQOOX. You can do this by manually turning the 

MAGNIFICATION 

knob or pressing the 

PRESET 

switch on the magnification display. 

8.  Locate area of interest using X and V stage shifts (you may need to go to low magnification 
see section I). 
9.

 

Turn the 

COARSE 

FOCUS knob until a nominally focused image is obtained.

 

10.

  

Press 

ABC 

button on the image adjust section to automatically correct brightness and 

contrast.

#

11. Press 

SMALL AREA DISPLAY 

button 

on scan speed section. 

12. Increase magnification to -10,000 times by manually turning the 

MAGNIFICATION 

knob. 
13. Focus using FINE 

FOCUS knob 

14. Refine sharpness of image with STIGMA/ALIGNMENT X AND Y knobs 
15. Return to 

TV 

or SLOW SCAN 

MODES 

on scan speed section. 

16. If you wish to move the sample to a different area, manipulate the specimen with the X 
AND Y shifts on the specimen chamber. You can also adjust the height, 

tilt, T and rotation, 

R of the sample. If you adjust the tilt or height of the sample, you will need to consider the 
effects of that movement with respect to any detectors inserted into the chamber. Always return 

X , Y, 

, R and T controls to their preset levels before removing a sample. 

11. When you have moved to a new area you may need to repeat steps 9-16. 
 
 
# The ABC control provides 'ideal brightness and contrast levels for photography. But may not always 
provide the levels of brightness and contrast you desire. An alternative method is to use the 
BRIGHTNESS and CONTRAST up/down arrowed keys on the image adjust section. 

Summary of Contents for McDATA Sphereon 4500

Page 1: ... switching DO NOT FORCE THEM 2 Turn the toggle switch SC AIRLOCK VALVE from CLOSED to AUTO OPEN a few seconds later the green light on the HV off ready display will come on If it doesn t please contact EM staff for help B INSERT SAMPLE 1 Put specimen stub onto specimen holder inc exact mechanism will vary depending upon the specimen type EM Unit staff will advise the best method for your sample wh...

Page 2: ...do this by manually turning the MAGNIFICATION knob or pressing the PRESET switch on the magnification display 8 Locate area of interest using X and V stage shifts you may need to go to low magnification see section I 9 Turn the COARSE FOCUS knob until a nominally focused image is obtained 10 Press ABC button on the image adjust section to automatically correct brightness and contrast 11 Press SMAL...

Page 3: ...ace sensitive information and reduce the size of the electron beam specimen interaction volume I Press FEll Vace 2 Using numeric keyboard key in desired accelerating voltage e g 5kV 3 Press ENTER 4 Microscope will automatically go to 5kV 5 Press PF16 EXIT to clear screen 6 Refocus image as described above 7 Go to ALIGN COLUMN section F F TO ALIGN THE COLUMN 1 Obtain a focused image at about 510 00...

Page 4: ...layed at INPUT will become active 5 Press PF16 EXIT to finish 6 Changing detectors may require you refocus the image and adjust brightness and contrast 7 n b The upper detector generally works better at short working distances 18 The lower detector generally works better at long working distances 18 See figure below You may accordingly need to change the specimen height H CHANGING THE WORKING DIST...

Page 5: ...urning the MAGNIFICATION knob or pressing the PRESET switch on the magnification display 3 Return X Y Z R and T controls to their preset levels listed on specimen chamber 4 Open airlock door by turning airlock level anticlockwise about one half turn DO NOT FORCE THIS LEVER it requires minimal effort to open 5 When airlock door is open fully insert specimen exchange rod into specimen chamber When r...

Page 6: ...ISHING A SESSION 1 Turn the SC AIRLOCK VALVE from AUTO OPEN to CLOSED the green light on the HV off ready display will turn off 2 Turn off the visual display by turning DISPLAY POWER to O 3 Fill in log book ...

Reviews: