28
‐
Q1E
‐
EP1430
8
Specifications
8.1
Probe
Type:
EUP
‐
V73W
Transvaginal
probe
Acoustic
working
frequency:
6.5MHz
Technology:
Convex
Array
Probe
Dimensions:
See
Fig.
21
Weight:
Approx.1kg
(Including
cable
and
connector)
Probe
materials:
Bio
‐
compatible
allergy
free
components
Acoustic
output:
According
to
IEC
60601
‐
2
‐
37
(See
Main
Unit
manual.)
Applicable
system:
Depending
on
production
and
upgrade
status.
For
detailed
information
contact
a
service
support.
Classification:
MDD
classification
IIa.
Cleaning:
Applicable
detergents
are
listed
in
the
suppliers
list
Disinfection:
Applicable
disinfectants
are
listed
in
the
suppliers
list
Sterilization:
ETO
gas
sterilization
or
Plasma
sterilization
Operating
conditions:
Ambient
temperature:
+10
–
+40
°
C
Contact
surface
temperature
(Temperature
of
examinee):
Max.
42
°
C
Relative
humidity:
30
–
75%
(Subject
to
no
condensation)
Storage/Transportation
conditions:
Temperature:
–
10
–
+50
°
C
Relative
humidity:
10
–
90%
(Subject
to
no
condensation)