-9-
Q1E-EP0385
3)
The immersible part of the probe should be left in the
detergent solution according to the specified contact time
of the detergent manufacturer.
4)
Wipe the un-immersible parts of the probe with a soft cloth
dipped with the detergent solution.
5)
Rinse the probe with running tap water for 1 minute.
(alternatively: immerse the immersible part of the probe in
a tray filled with deionized water/tap water (see Fig.3) for
5 min.)
6)
Visually check the outer surface of the probe for cleanness.
If necessary, use magnifying glass for visually check. If
there is still soil visible, repeat all above steps.
Water
Detergent
Disinfectant
Fig.3 Immersion of the probe