-19-
Q1E-EP1359
B)
Magnetic sensor attachment
1)
Prepare the disinfectant solution as stated in the procedure for the
probe.
2)
Immerse the Magnetic sensor attachment into the disinfectant (see Fig.
8). Set a clock to insure the recommended contact time which is 5
minutes.
3)
Rinse the Magnetic sensor attachment with deionized water for 1 minute.
(alternatively: immerse it in a tray filled with deionized water (see
Fig. 8) for 5 min.)
4)
Visually check the outer surface of the Magnetic sensor attachment
for leavings of the disinfectant. If necessary, repeat the rinsing.
Fig. 8 Immersion of the probe and the Magnetic sensor attachment
Water
Detergent
Disinfectant