- 5 -
■
Operation procedure
●
How to check torque using Fidaptor
First, connect the torque meter (HM) to the detector.
Then turn on the power and confirm if the battery has been charged.
・
How to check the battery
①
Turn on the power.
②
Set the measurement mode to TRACK.
③
If the battery is low, “LOBAT” is displayed in the upper left corner of the display.
Then, charge the battery (do not charge longer than eight hours). Use the chargers for this battery.
(Note) Do not take measurements while charging the battery.
2. Fix the detector. There are two points for fixing. (See the dimensional drawing of the detector on
page 15.)
3. Connect the cord to the detector.
Connect the cord to the detector. There are two connectors. Use one of them considering the
workability.
*Make sure the connector has the correct shape. If you ignore this, it may cause pin bending.
(Note)
There are two connectors.
Use one of them and cover the other one with a cap.
Points for fixing
4. Use the handy screwdriver to loosen the threaded shaft ring of Fidaptor a little.
A
spring
loosens
Loosen the shaft so that you can turn the spring with your fingers.
The shaft should always go back to the same position.
5. Set Fidaptor to the socket.
5-1 Fix Fidaptor to the socket.
LOBAT
When “LOBAT” is displayed, start
charging the battery.
●
Important
In order to fix Fidaptor, use the L-shaped wrench
(1.5mm) to fasten four hollow set screws (M3.0) on
the side of the socket with even torque.
This way you can measure in a stable manner.
Hollow set screw (M3, 4 pieces)
L-shaped wrench, 1.5mm
Summary of Contents for HM-10
Page 20: ......