226
Error Display
15.3 Error Display
When an error is displayed in the LCD display area, the instrument requires a check or repair. Please
contact your authorized Hioki distributor or reseller.
Errors by the protection function
Error display
Description
Ext. I/O output
Solution
POWER SUPPLY ERROR
An error in the power supply
circuit has been detected.
PROTECTION
The instrument is malfunctioning. Turn
off the instrument and make a repair
request.
FAN ERROR
A fan stop has been detected.
PROTECTION
Turn off the instrument immediately and
check the instrument installation and fan
condition. If the situation still does not
improve, there is a risk of malfunction.
Make a repair request.
LED ERROR
DANGER lamp LED failure has
been detected.
PROTECTION
Turn off the instrument and make a repair
request.
INTERLOCK STATE
Input of an interlock signal has
been detected and the interlock
has been activated.
PROTECTION
Release the interlock (p. 165).
REMOTE CONNECTOR
Connection/disconnection of
the external switch terminal has
been detected.
PROTECTION
Check the connection of the Remote
Control Box and press the
STOP
button
to clear the error.
GROUND FAULT
The GFI function has been
activated (p. 112).
PROTECTION
Check the connection and safety of the
test target.
OVERLOAD
A current three times the
maximum rated current has
been detected.
PROTECTION
Check the connection and the test target
and press the
STOP
button to clear the
error.
SUPPLY VOLTAGE FREQ
ERROR
The main power frequency is
outside the specifications.
PROTECTION
Check the power supply.
POWER TEMP ERROR
A temperature error in the
power supply circuit has been
detected.
PROTECTION
Turn off the instrument immediately and
stop using the instrument. If the situation
still does not improve, there is a risk of
malfunction. Make a repair request.
SUPPLY VOLTAGE
ERROR
The output voltage of the power
supply circuit has dropped.
PROTECTION
Check for an instantaneous power
outage with the power supply. Press the
STOP
button to clear the error.
HIGHLOAD POWER
OUTAGE
A relatively longer voltage dip
occurs during voltage output
when the load is large.
PROTECTION
OVERLOAD POWER
OUTAGE
PROTECTION
OUTPUT TIME LIMIT
The test time has exceeded the
rated time.
PROTECTION
Perform a text in the protection time
rating (p. 58). Press the
STOP
button
to cancel.
HIOKI ST5680A961-00
Summary of Contents for ST5680
Page 2: ...HIOKI ST5680A961 00 ...
Page 84: ...84 Individually Executing Specific Steps HIOKI ST5680A961 00 ...
Page 118: ...118 Key Lock HIOKI ST5680A961 00 ...
Page 154: ...154 Calibration Expiration Check Function HIOKI ST5680A961 00 ...
Page 182: ...182 Internal Circuit Configuration HIOKI ST5680A961 00 ...
Page 220: ...220 Option Specifications HIOKI ST5680A961 00 ...
Page 238: ...238 External View 16 4 External View Unit mm HIOKI ST5680A961 00 ...
Page 250: ...250 HIOKI ST5680A961 00 ...
Page 253: ...www hioki com HIOKI ST5680A961 00 ...