58
Starting and Ending a Test
4.15 Starting and Ending a Test
Output voltage time rating
CAUTION
Use the instrument within the time rating in the table below.
When the instrument is to be used consecutively, the output time is limited by the
test voltage. The maximum test time and standby time vary depending on the test
voltage. If the instrument is used over the time rating, it could be damaged due to
heating.
0
0
1 kV 2 kV 3 kV 4 kV 5 kV 6 kV 7 kV 8 kV
50 mA
20 mA
100 mA
1
M
Ω
10
k
Ω
100
k
Ω
40
k
Ω
1000 V
2000 V
Load current
Lower reference value
Test voltage
DC withstand voltage
test mode
Insulation resistance
test mode
Test voltage
2 min.
5 min.
Continuous
Continuous
15 min.
5 min.
30 min.
30 min.
15 min.
15 min.
If a test continues for the maximum test time, leave the instrument for 30 minutes.