44
Judging Measured Values (Comparator function)
3.5 Judging Measured Values (Comparator
function)
The following are enabled with the comparator function.
•
The instrument displays a judgment result (COMP lamp U.FAIL/PASS/L.FAIL).
U.FAIL Upper limit value ≤ Measured value
PASS
Within judgment standard
L.FAIL Lower limit value ≥ Measured value
•
A buzzer sounds (default value FAIL).
Refer to “3.7 Announcing the Judgment Results Using a Beep Sound” (p. 47).
•
A judgment result is externally output.
Refer to “7 External Control (EXT.I/O)” (p. 81).
There are the following three comparator judgment methods.
Upper/lower limit value setting range: 0.000 to 4000 M
Ω
Judgment by the
lower limit value
Example: 100 M
Ω
PASS
LOWER FAIL
(L.FAIL)
Lower limit
value [M
Ω
]
Judgment by the
upper limit value
Example: 100 M
Ω
PASS
UPPER FAIL
(U.FAIL)
Upper limit
value [M
Ω
]
Judgment by the
upper and lower limit
values
Example: 1000 M
Ω
Example: 100 M
Ω
Lower limit
value [M
Ω
]
LOWER FAIL
(L.FAIL)
PASS
UPPER FAIL
(U.FAIL)
Upper limit
value [M
Ω
]
If the upper and lower limit values are set to values outside of the accuracy range,
the U.FAIL and L.FAIL lamps turn on at the same time.
Summary of Contents for ST5520
Page 2: ......
Page 8: ...vi Contents ...
Page 22: ...14 Usage Notes ...
Page 56: ...48 Announcing the Judgment Results Using a Beep Sound ...
Page 82: ...74 Initializing the System Reset ...
Page 88: ...80 Deleting the Panel Data ...
Page 158: ...150 Command Reference ...
Page 193: ...Appx 13 Rack Mounting Screw location dimensional drawing 10 9 8 7 6 5 4 3 2 1 Appx Ind ...
Page 194: ...Appx 14 Dimensional Diagram Appx 8 Dimensional Diagram 166 2 215 1 14 1 80 1 ...
Page 197: ...13 09 ...
Page 198: ......
Page 199: ......
Page 200: ......
Page 201: ......