106
Double Action Function
8.3
Double Action Function
The double action function prevents misoperations during test startup.
When the double action function is enabled, merely pressing the
START
button does not start the
test. Press the
STOP
button and then press the
START
button within approx. 1 sec. to start the
test.
(Measurement screen)
[SYSTEM]
>
[SYSTEM]
1
1
Tap
[DBL ACTN]
.
OFF
The double action function is enabled.
ON
The double action function is disabled.
IMPORTANT
When sending trigger signals using communications and external control (EXT. I/O), testing will start even if
the double action function is enabled and the
STOP
button is not pressed.