3.5 Checking Measured Values
52
When measuring close to 0
Ω
, measured values may turn negative. If measured values turn
negative otherwise, check the following:
•
Are the SOURCE or SENSE wires connected backwards?
→
Rewire correctly.
•
Has the contact resistance decreased since you performed zero-adjustment?
→
Repeat the zero-adjustment process.
•
Is the scaling calculation result negative?
→
Change the scaling settings.
You can change what information is shown on the Measurement screen.
Displaying temperature and pre-calculation measured values
3.5
Checking Measured Values
The resistance value will be displayed.
•
If the display does not indicate the mea-
sured value, see "Confirming Measure-
ment Faults" (p. 55).
•
To convert the value into a parameter
other than resistance, see below.
See:
"5.4 Performing Temperature Rise Test
(Temperature Conversion Function (
See:
Switching the Display
[VIEW]
Switch the Measurement
screen.
Switch the function menu
to P.1/2.
1
2
You can switch this part of the display to
show nothing, the temperature, or the pre-
calculation measured value.
See:
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......