4.6 Correcting Measured Values and Displaying Physical Properties Other than Resistance
77
4
This function applies a correction to measured values. It can be used to cancel the effects
of the probing position or differences between measuring instruments, or to apply a user-
specified offset as an alternative to zero-adjustment. In addition, units can be specified,
allowing it to be used to convert measured values to physical properties other than resis-
tance (for example, length).
Scaling is performed by means of the following equations:
R
S
=
A
×
R
+
B
R
S
: Resistance value after scaling
R
: Measured value after zero-adjustment and temperature correction
A
: Gain coefficient Setting range: 0.2000 × 10
-3
to 1.9999 × 10
3
B
: Offset
Setting range: 0 to ±9 × 10
9
(maximum resolution: 1 n
Ω
)
Displayed and sent/received measured values as well as the printer output format vary with
the gain coefficient.
4.6
Correcting Measured Values and Displaying
Physical Properties Other than Resistance
Values (Scaling Function)
Low-Power: OFF
* When high-precision mode is off in the 100 M
Ω
range, 4 digits are displayed.
Low-Power: ON
Range
Gain coefficient
(0.2000 to
1.9999)
×10
-3
(0.2000 to
1.9999)
×10
-2
(0.2000 to
1.9999)
×10
-1
(0.2000 to
1.9999)
×1(10
0
)
(0.2000 to
1.9999)
×10(10
1
)
(0.2000 to
1.9999)
×10
2
(0.2000 to
1.9999)
×10
3
10 m
Ω
00.000
μ
000.000
μ
0000.000
μ
00.000 00 m
000.000 0 m
0000.000
m 00.000 00
100 m
Ω
000.000
μ
0000.000
μ
00.000 00 m 000.000 0 m
0000.000
m
00.000 00
000.000 0
1000 m
Ω
0000.000
μ
00.000 00 m 000.000 0 m 0000.000
m
00.000 00
000.000 0
0000.000
10
Ω
00.00 000 m 000.000 0 m 0000.000
m 00.000 00
000.000 0
0000.000
00.000 00 k
100
Ω
000.000 0 m 0000.000
m 00.000 00
000.0000
0000.000
00.000 00 k
000.000 0 k
1000
Ω
0000.000
m 00.000 00
000.000 0
0000.000
00.000 00 k
000.000 0 k
0000.000
k
10 k
Ω
00.000 00
000.000 0
0000.000
00.000 00 k
000.000 0 k
0000.000
k
00.000 00 M
100 k
Ω
000.000 0
0000.000
00.000 00 k
000.000 0 k
0000.000
k
00.000 00 M 000.000 0 M
1000 k
Ω
0000.000
00.000 00 k
000.000 0 k
0000.000
k
00.000 00 M
000.000 0 M 0000.000
M
10 M
Ω
00.000 00 k
000.000 0 k
0000.000
k
00.000 00 M
000.000 0 M
0000.000
M 00.000 00 G
100 M
Ω
* 000.000 0 k
0000.000
k
00.000 00 M 000.000 0 M
0000.000
M
00.000 00 G 000.000 0 G
1000 M
Ω
0000.0
k
00.000
M 000.00
M 0000.0
M
00.000
G
000.00
G 0000.0
G
Range
Gain coefficient
(0.2000 to
1.9999)
×10
-3
(0.2000 to
1.9999)
×10
-2
(0.2000 to
1.9999)
×10
-1
(0.2000 to
1.9999)
×1(10
0
)
(0.2000 to
1.9999)
×10(10
1
)
(0.2000 to
1.9999)
×10
2
(0.2000 to
1.9999)
×10
3
1000 m
Ω
0000.00
μ
00.000 0
m 000.000
m 0000.00
m
00.000 0
000.000
0000.00
10
Ω
00.000 0
m 000.000
m 0000.00
m 00.000 0
000.000
0000.00
00.000 0
k
100
Ω
000.000
m 0000.00
m 00.000 0
000.000
0000.00
00.000 0
k
000.000
k
1000
Ω
0000.00
m 00.000 0
000.000
0000.00
00.000 0
k
000.000
k
0000.00
k
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......