4.13 Increasing the Precision of the 100 M
Ω
Range (100 M
Ω
High-precision Mode)
96
The precision of the 100 M
Ω
range can be increased.
Turning on high-precision mode has the following effects:
•
The 1,000 M
Ω
range will be unavailable for use.
•
More time will be required for measured values to stabilize. To adjust the time required
until values stabilize, set a delay.
See:
"4.9 Setting Pre-Measurement Delay" (p.84)
4.13 Increasing the Precision of the 100 M
Ω
Range (100 M
Ω
High-precision Mode)
1
Open the Settings Screen.
2
Open the Measurement Setting Screen.
3
Turn on 100 m
Ω
range high-precision mode.
4
Return to the Measurement screen.
The Settings screen
appears.
Switch the function menu
to P.2/3.
1
2
Move the cursor to the
[MEAS] tab with the left and
right cursor keys.
2
Enables the 100 M
Ω
high-
precision mode
Disables the 100 M
Ω
high-
precision mode
Selection
1
Return to the
Measurement screen.