4.9 Setting Pre-Measurement Delay
84
This function adjusts the time for measurement to stabilize by inserting a waiting period
after use of the OVC or the auto range function to change the measurement current. When
this function is used, the instrument waits for its internal circuitry to stabilize before starting
measurement, even if the measurement target has a high reactance component.
If the measurement target, for example, is an inductor that takes longer to stabilize after
applying a measurement current, and it cannot be measured with the initial delay (default),
adjust the delay. Set the delay time to approximately ten times the following calculation so
that the reactance component (inductance or capacitance) does not affect the measure-
ment.
The delay setting can be selected from a preset (internal fixed value) or user-set value.
(1) Preset (internal fixed value)
Value depends on the range and offset voltage correction function.
(2) User-set value
Setting range: 0 to 9999 ms
The set value is used for all ranges.
4.9
Setting Pre-Measurement Delay
L
...... Inductance of measurement target
R
...... Resistance of measurement test leads + contacts
I
....... Measurement current (see “Accuracy” (p.251))
V
O
... Open-terminal voltage (see “Accuracy” (p.251))
L
R
t =
−
ln 1
−
I R
V
O
LP ON
Delay
1
LP OFF (unit: ms)
Range
100 M
Ω
range
high-precision
mode
Measure-
ment cur-
rent
Delay
OVC: OFF OVC: ON
10 m
Ω
−
−
75
25
100 m
Ω
−
High
250
25
−
Low
20
2
1000 m
Ω
−
High
50
2
−
Low
5
2
10
Ω
−
High
20
2
−
Low
5
2
100
Ω
−
High
170
2
−
Low
20
2
1000
Ω
−
−
170
2
10 k
Ω
−
−
180
−
100 k
Ω
−
−
95
−
1000 k
Ω
−
−
10
−
10 M
Ω
−
−
1
−
100 M
Ω
ON
−
500
−
100 M
Ω
OFF
−
1
−
1000 M
Ω
OFF
−
1
−
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......