2.3 Turning the Power On and Off
25
2
Turning Power On
When the power is turned on, the same setting as when the power was last turned off
appears (backup function).
When powered up for the first time, the default settings appear.
See:
Before Starting Measurement
To obtain precise measurements, provide about 30 minutes warm-up after turning power on.
Measurement settings are recalled from when the power was previously turned off (settings
backup).
However, measurement settings made through the RS-232C or GP-IB interface are not
retained, although they can be stored using the
:SYSTem:BACKup
command (p. 135).
Turning Power Off
Press the
POWER
button (it lights red when the instrument is off).
Disconnect the power cord from the outlet to extinguish the
POWER
button light.
When power is turned on again, operation resumes with the same settings as when last
turned off.
2.3
Turning the Power On and Off
Press the
POWER
button (it lights green).
Self-test
Indicates an error (p. 181).
After Power-On
A self-test (instrument diagnostic routine) is performed.
During the self-test, the following information is displayed while the hardware is verified.
Error
No Errors
Normal display (measurement screen)
The following information is displayed during self-testing:
•
Manufacturer and model name
•
Firmware versions (main, and measurement)
•
Communication setting
•
Line frequency setting
If a power outage (e.g., breaker trip) occurs when the instrument is on, it will automatically
turn on again when power is restored (without pressing the
POWER
button).
Summary of Contents for RM3542
Page 2: ......
Page 6: ...Contents iv ...
Page 16: ...Operating Precautions 10 ...
Page 26: ...1 3 Screen Organization 20 ...
Page 32: ...2 3 Turning the Power On and Off 26 ...
Page 46: ...3 8 Confirming Faulty Measurements 40 ...
Page 64: ...4 12 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 58 ...
Page 84: ...6 4 Auto Exporting Measured Values at End of Measurement Data Output Function 78 ...
Page 90: ...7 3 Printing 84 ...
Page 172: ...9 10 Device Compliance Statement 166 ...
Page 190: ...11 4 Disposing of the Instrument 184 ...
Page 200: ...Appendix 5 Dimensional Diagram A10 Appendix 5 Dimensional Diagram ...
Page 214: ...Index Index 4 ...
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......