3.1 Pre-Operation Inspection
27
3
Refer to "Measurement Flow" (p. 3) for an outline of the measurement process from preparation to
end-of-measurement.
Measurement Settings
(Basic Measurements)
Chapter 3
3.1 Pre-Operation Inspection
Do not use the instrument if damage is
found, as electric shock or short-circuit
accidents could result.
Contact your dealer or Hioki represen-
tative.
Metal Exposed
Is the power cord insulation torn, or is any
metal exposed?
1
No Metal Exposed
Before using the instrument for the first time, verify that it operates normally to ensure that
no damage occurred during storage or shipping. If you find any damage, contact your dealer
or Hioki representative.
Peripheral Device Inspection
Is the insulation on a measurement probe
torn, or is any metal exposed?
Metal Exposed
If there is any damage, measurement
values may be unstable and measure-
ment errors may occur.
Replace the cable with an undamaged
one.
No Metal Exposed
If damage is evident, request repairs.
Yes
Is damage to the instrument evident?
Instrument Inspection
When turning power on
Does the self-test screen appear (model
name, version no.) (p. 25)?
Not displayed
2
The power cord may be damaged, or
the instrument may be damaged inter-
nally. Request repairs.
Does the Measurement screen appear after
self-test?
No
Displayed
The instrument may be damaged inter-
nally. Request repairs.
See:
"11.1 Troubleshooting" (p. 195)
"11.3 Error Displays and Solu-
tions" (p. 197)
An error is
displayed (
Err
)
Displayed
Inspection complete
Summary of Contents for RM3542-50
Page 1: ......
Page 2: ......
Page 6: ...Table of Contents iv...
Page 26: ...1 3 Screen Organization 20...
Page 32: ...2 3 Turning the Power On and Off 26...
Page 48: ...3 8 Confirming Faulty Measurements 42...
Page 72: ...4 14 Compensating for Thermal EMF Offset Offset Voltage Compensation OVC 66...
Page 84: ...5 7 Initializing Reset 78...
Page 94: ...6 4 Auto Exporting Measurement Values at End of Measurement Data Output Function 88...
Page 206: ...11 4 Disposing of the Instrument 200...
Page 216: ...Appendix 5 Dimensional diagram A10 Appendix 5 Dimensional diagram Unit mm...
Page 230: ...Index Index 4...
Page 231: ...HIOKI E E CORPORATION...
Page 232: ......
Page 233: ......
Page 234: ......