PW3365-20
CLAMP ON POWER LOGGER
Measurement Guide
Apr. 2018 Revised edition 1 Printed in Japan
PW3365A984-01 18-04H
Thank you for purchasing the HIOKI PW3365-20
Clamp On Power Logger.
This guide introduces the PW3365-20's basic
measurement procedure with the Quick Set to
first time users.
Before using the instrument
,
be sure
to read
the Instruction manual carefully.
Easy configuration with the
Set
How to configure electric energy measurement for a
3-phase 4-wire 220 V line
Setting Items
Setting Example
Wiring
: 3P4W (3-phase 4-wire)
Clamp sensor
: Model 9661 (500 A rating)
Current range
: 50 A
Save to...
: SD memory card
Save interval
: 5 minutes
Save items
: Average only
Folder/Filename
: Automatic
Rec. start method
: Interval
Rec. stop method
: Manual
Clock setting
:
User-specified
Measurement frequency : 50 Hz
You will need
Model PW3365-20
Model PW9020 Safety Voltage Sensor x4
Model 9661 Clamp on
Sensor (optional) x 3
Model Z1008 AC Adapter
SD memory card
(optional)
Color clips for clamp
sensors
Red (CH1)
Yellow (CH2)
Color clips for
voltage sensors
Red (CH1)
Yellow (CH2)
Concept image
of measurement
3-phase 4-wire
220 V line
Quick
Blue (CH3)
Blue (CH3)
Preparations
1
Attach the color clips.
(CH1)
(CH2)
(CH1) (CH2)
(N)
2
Insert the SD memory
card. (on right side of
instrument)
3
Connect the AC adapter.
(
on left side of instrument
)
Be sure to provide a Hioki
optional SD Memory Card.
Operation is not guaranteed
with other SD memory cards.
Model 9661
Model PW9020
Red
Yellow
Red
Yellow
None
(CH3)
Blue
(CH3)
Blue
1. Starting the Quick Set
1
Turn on the instrument.
(
on left side of instrument
)
2
Press the
key.
The Quick Set Start dialog will
be displayed.
First time powered
on only
Language setting
Measurement
frequency setting
(50 Hz)
3
Press the
Enter
key.
2. Basic settings
3P4W
9661 (500
A)
SD CARD
Set the current time.
1
Configure settings as
shown in screenshot
below.
2
Press the
F2 [NEXT]
key.
2
3
1
Voltage sensor input terminals
(Connect voltage sensors.)
Left
Upper
POWER Switch
Current sensor input terminals
(Connect clamp sensors.)
AC adapter
connection jack
Right
SD memory
card slot
Names of Parts (excerpt)
3. Connect the sensors to the instrument.
1
Connect the voltage sensors
to the voltage sensor input
terminals.
2
Connect the clamp sensors
to the current sensor input
terminals.
3
Be sure that the SD memory
card is inserted. (on right
side of instrument)
4
Press the
F2 [NEXT]
key.
Clamp sensor connector (BNC)
Current sensor
input terminals
Line up
Line up
Lock the
connector.
1
Insert
2
Match the color of each sensor’s color clip to the color
of the terminal.
Voltage sensor connector
Voltage sensor input terminals
Align
Insert
4. Connecting voltage sensors to the measurement target
1
Refer to the wiring diagram to check
the locations to which you have
connected the voltage sensors.
2
Connect the voltage sensors to the
secondary side of the breaker.
3
Check the readings.
In this example
,
the screen should indicate
approximately 220 V and 50 Hz if there is
no problem.
(Because the measurement target is a 220
V line
,
and the frequency being measured
is 50 Hz.)
4
Verify the results of checking the
wiring.
5
Press the
F2 [NEXT]
key.
All results are
.
Move the cursor to
item.
Press the
Enter
key.
Check the contents of the dialog box
and correct the wiring.
If there is a
result
1
2
3
Tip
The power source side of the breaker is called the primary
side; whereas the load side
,
the secondary side.
For your safety
,
connect the voltage sensors and the clamp
sensors to the secondary side.
Primary side
Source side
Secondary side
Load side
Breaker
Yellow
Proper application
C
A
B
Align the insulated wire with the
marks on the voltage sensor to
the wire.
Red
None
For more information
,
see “3.6 Connecting
the Voltage Sensors to Target to be
Measured” in the instruction manual.
Improper application
Failure to apply the sensor properly will prevent you from being to
make an accurate measurement.
Example: 3-phase 4-wire 220 V line
Secondary side of breaker
Clamped with the
tips of the clip
Clamped too far
back in the clip
Clamped with the
measurement target at an
angle
Clamping targets with
different voltages at the
same time
Secondary
side
(Load side)
A
B
C
OK
Blue
N
EN