Specifications
(Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
High-speed function
(high speed recording)
Time axis
100μs to 100ms/div, 10 range, resolution: 100 points/div
Sampling period
1/100 of time axis ranges
(minimum sampling period 1 μs, all channels simultaneously)
Recording length
5 to 10000 divisions fixed
(5division steps)
Automatic save func-
tion
Binary data, text data, calculation results, calculation results,
text + calculation results, or NONE
Other save functions
Save and delete function: ON/OFF
Screen settings
Split screen
(1, 2, or 4 segments)
, X-Y waveform compositing
(1 screen)
Pre-trigger
Can record data from before the trigger point, 0 to 100 % of
recording length; 13 settings, or user-configured
Waveform scrolling
Backwards scrolling through past waveform data both during and
after measurement
Calculation functions
Up to four arithmetic operations simultaneously
Average value, effective (RMS) value, peak to peak value,
maximum value, time to maximum value, minimum value, time to
minimum value, period, and frequency, area, X-Y area.
Real-time function
(actual time recording)
Recording interval
100μs to 500μs, 1ms to 500ms, 1s to 1min, 19 settings
Display time axis: 10ms to 1day/div, 22 ranges
Real-time printing
(with optional MR9000)
ON/OFF
*Simultaneous printing: Supported when using a time axis slower than 1 s/div.
Recording Time
Continuous save to CF card or USB memory can be set ON/OFF
Envelope mode
ON/OFF
Waveform recording
The last 1 Mwords
(before measurement was stopped)
are saved in
the instrument’s internal memory
(when envelope mode is on, 500
kwords).
Real-time save func-
tion
Binary data, text data, calculation results, calculation
results, text + calculation results, or NONE
Other save functions
Split save: ON/OFF/fixed time
Save and delete: ON/OFF
Eject media: Media can be ejected while saving data in real time.
Event marks
1) Event marks can be input during measurement (up to 100 marks).
2) Can move to waveform before or after an event mark based on
specified event number input.
Trigger function
Repeat recording
Single/Repeat
Trigger timing
High-speed function: Start
Real-time function: Start, Stop, Start & Stop
Trigger conditions
AND/OR supported for all trigger sources
Trigger source
Trigger sources can be selected for each channel. Instrument enters
free-run mode when all trigger sources are off.
1) Analog input CH1 - CH4
2) Logic input LA1 - LA4, LB1 - LB4 (4ch × 2 probes)
3) External trigger
4) Interval trigger: Fixed-time recording for specified measurement
interval (month/day/hours/minutes/seconds)
Trigger types
1) Level
2) In
3) Out
4) Voltage drop
(High-speed function) : For AC 50/60 Hz power lines
5) Waveform judgment
(High-speed function): For AC 50/60 Hz power lines
6) Logic
7) External: Rising edge/falling edge
Level setting resolution
0.1 % f.s.
(f.s.=10 div)
Trigger filter
High-speed function: 7 settings from 10 to 1000 samples or OFF
Real-time function: ON/OFF
Trigger output
Open collector (5 V output, active Low)
Analog input
(Accuracy defined at 23° ±5°C, 80% rh or less, for measurements taken following
zero adjustment 30 minutes after instrument is turned on)
Measurement
functions
4-channel voltage measurement; switchable between instantaneous
value (waveform) and RMS value
Input connectors
Isolated BNC connector
(input impedance 1 MΩ, input capacitance 7 pF)
Max. rated voltage to
earth
600 V AC, DC CAT III / 300 V AC, DC CAT IV
(with input isolated from the unit, the maximum voltage that can be applied between
input channel and chassis and between input channels without damage)
Measurement range
10 mV to 100 V/div, 13 ranges, full scale: 10 div, AC voltage that can
be measured and displayed using high-speed function: 600 Vrms
Low-pass filter: 5 Hz/50 Hz/500 Hz/5 kHz/50 kHz
Measurement resolu-
tion
1/640 of measurement range
(using 14-bit A/D conversion, at × 1)
Highest sampling rate
1 MS/s
(simultaneous sampling in 4 channels)
Instantaneous value
measurement accuracy
±0.5% f.s. (after zero-adjust)
RMS measurement
RMS accuracy: ±1.5% f.s.
(30Hz to 1kHz)
±3% f.s. (
1kHz to 10kHz)
Response time: 300ms
(rising edge 0 to 90% of full scale with filter off)
Crest factor: 2
Frequency charac-
teristics
DC to 100 kHz ±3dB
Input coupling
DC/GND
Max. rated voltage
between terminals
600 V AC, DC
(maximum voltage which when applied to between input
terminals does not damage them)
Basic specifications
Measurement func-
tions
High-speed function (high speed recording)
Real-time function (actual time recording)
Number of channels
4 8 logic
Isolated analog channels, isolated input and outputs, logic has common GND.
Maximum sampling rate
1Msamples/s (1 μs cycle, all channels simultaneously)
Memory capacity
14bit × 1 M words/ch (1 word = 2 bytes, not expandible)
External memory
CF card slot × 1 (Up to 2 GB, supports FAT16 and FAT32 formats)
USB memory × 1 (USB 2.0 -A receptacle)
Time accuracy
(at 23°C)
Sampling time accuracy: ±0.0005 %, Clock precision: ±3s/day
Backup function
(reference value at
23°C)
Clock and settings: 10 years or more (at 25°C / 77°F)
Waveform backup function: Approx. 40 minutes
•
When instrument is powered off at least 3 minutes after being turned on
External control
External trigger input, Trigger output, external start input, external
stop input, status output, ground pin
Interface
USB: 1 port USB 2.0 High Speed mini-B receptacle
Functions: Configure settings/perform measurement using communications
commands: transfer file stored in CF/USB memory to computer (USB drive
mode)
Environmental condi-
tions for use
(no condensation)
Temperature range: -10°C (14°F) to 50°C (122°F)
Humidity range: -10°C (14°F) to 40°C (104°F), 80% rh or less
40°C (104°F) to 45°C (113°F), 60% rh or less
45°C (113°F) to 50°C (122°F), 50% rh or less
When powered by BATTERY PACK Z1000:
0°C (32°F) to 40°C (104°F), 80% rh or less
When recharging the Z1000: 10°C (50°F) to 40°C (104°F), 80% rh or less
Environmental condi-
tions for storage
(no condensation)
Temperature range: -20°C (-4°F) to 60°C (140°F)
Humidity range: 80% rh or less (-20°C (-4°F) to 40°C (104°F)), 60% rh
or less
(40°C (104°F) to 45°C (113°F)), 50% rh or less (45°C (113°F) to 60°C
(140°F))
BATTERY PACK Z1000: -20°C (-4°F) to 40°C (104°F) , 80% rh or less
Compliance
standard
Safety: EN61010 EMC: EN61326, EN61000-3-2, EN61000-3-3
Vibration resistance: JIS D 1601, Type 1: passenger vehicle,
Conditions: equivalent to Type A
Power requirements
Note: LR6/AA alkaline batteries
are not sufficient to power the
unit when it is connected with
the Printer Unit MR9000. Use of
other power supplies is required.
(Continuous operating time is
given as a reference value at 23°C.)
1)
AC ADAPTER Z1002
: 100 to 240V AC (50/60 Hz)
2)
BATTERY PACK Z1000
: 7.2V DC
Continuous operating time: Approx. 3 hours with backlight on, approx.
3.5 hours with backlight off
(AC adapter has priority when both are used)
3) LR6 (AA)×8
Approx. 40 minutes with backlight on. Approx. 50minutes with back-
light off.
(when used with AC adapter, AC adapter takes precedence)
4) 10 to 28V DC (using special order cable)
Charging functions
(reference value at 23°C)
Charging time is about 3 hours
(can be charged by connecting the AC adapter while the Z1000 battery
pack is attached)
Max. rated power
1) When instrument is powered with the Z1002 AC adapter or an
external DC power supply: 11 VA*
1
, 10 VA*
2
, 40 VA*
3
2) When instrument is powered with the Z1000 battery pack;
9 VA*
1
, 8 VA*
2
, 22 VA*
3
*
1
Real-time data storage, backlight on
*
2
Real-time data storage, backlight off
*
3
Real-time data storage, backlight on, with printer used
Dimensions, mass
(including battery pack)
205 mm (8.07 in)W × 199 mm (7.83 in)H × 67 mm (2.64 in)D, 1.66 kg
(58.6 oz) (printer detached)
303 mm (11.93 in)W × 199 mm (7.83 in)H × 67 mm (2.64 in)D,2.16 kg
(76.2 oz) (printer attached)
Accessories
Instruction manual ×1, AC adapter Z1002 ×1, Alkaline battery box
×1, Strap ×1, USB cable ×1, Application disk (Wave viewer Wv,
Communication commands table) ×1
Function
Presets
Select from basic measurement guide, example measurement guide,
and commands for loading internally stored settings.
Scaling function
Select decimal or scientific notation for each channel.
1) Scaling ratio: Select scaling ratio, offset value, and units.
2) Two-point configuration: Set input values, post-scaling values,
and units.
3) HIOKI sensor: Set HIOKI clamp-on probe and range value.
4) Output rate setting: Select scaled value per 1 V from a list.
Data protection
Open files are closed before the instrument turns itself off when a
power outage occurs while saving data to recording media. When
powering the instrument with a battery, open files are closed and access
to the media is stopped when remaining battery power falls below a
certain level.
*Valid when at least 3 minutes has elapsed since the instrument was turned on.
Reservation function
Up to 10 measurement start and measurement stop conditions can be set.
Other
Settings can be automatically loaded from internal memory or
media when the instrument is turned on. Up to 10 settings can be
saved in the instrument’s internal memory.
Printer
(Printer Unit MR9000 docks onto the main device)
Features
Printer paper one-touch loading, high-speed thermal printing
Printer paper
112 mm (4.4 in) × 18 m (59.06 ft), thermal paper roll (using 9234)
Recording width: 100 mm, 10 div f.s., 1 div=10 mm (80 dot/div)
Recording speed
Max. 10 mm/s (0.39 inch/s)
(Printing is not supported when using alkaline batteries.)