Hioki MR8847A Manual Download Page 4

4

An Extensive Line of Units 

for Detecting a Wide Range 

of Phenomena

Combine multiple units to record a range of phenomena 

simultaneously.  For  example,  use  five 

4CH Analog Units

 to 

measure 20 analog channels and three 

Logic Units

 to measure 

up to 64 channels relay on/off signals or PLC (programmable 

logic controller) signals. That’s simultaneous measurement of 

84 channels!

Simultaneous measurement of multiple 

locations across 32 channels at 5 MS/s

Simultaneously measure up to 32 channels

The U8975 accepts direct input of up to 200 V DC across 4 channels. 

With a sampling rate of 5 MHz (across a frequency band of 2 MHz), 

high speed, and 16-bit resolution, it can perform multi-channel, 

high-speed, and high-resolution measurement.  

4CH Analog Unit U8975 

NEW

700 V AC

1000 V DC

4 ch

200 V DC

4 ch

100 mV f.s.

3275

(0.01 V/A)

(10 V/A, 1 V/A, 0.1 V/A)

CT6711

3 mV

4 mV

2 mV
1 mV

40

0

80

120

160

Elapsed time [ms]

During measurement with the CT6711 (10 V/A range)

Maximum 

resolution

0.1 μV

Battery

Battery package

Direct, high-voltage input without differential probes

The U8974 is ideal for measuring the primary and secondary sides of UPS 

power supplies and commercial power supply transformers. It can measure 

high-voltage power lines, including 380 V and 480 V circuits found in many 

countries. With high-speed sampling at up to 1 MS/s and 16-bit resolution, it 

can also be used in load rejection testing and switch testing. 

High Voltage Unit U8974

Analyze correlations between phenomena, including voltage levels before 

and after generator disconnection, RPM fluctuation rates, governor servo 

operating status, and voltage governor switching timing. 

Simultaneously measure up to 32 channels 

at high resolution 

Thanks to four input channels and a high-sensitivity 100 mV f.s. range, the 

U8978 can measure multiple channels of output from a variety of sensors. The 

unit is ideal for use in measuring currents of various magnitudes in the 

development of automobile accessory controls. Utilized in combination with the 

multi-range Current Probe CT6711, it can measure currents from 1 mA to 50 A. 

4CH Analog Unit U8978

NEW

Observe minuscule currents using high-sensitivity 

wideband current probes

Current probe lineup

Analyze minuscule current waveforms 

from low-power-consumption devices in 

100 μA resolution. Record device current 

consumption waveforms in high resolution 

over extended periods of time. 

Current consumption waveform for a temperature and humidity sensor

Specifically designed for DC voltage measurement 

with extremely high precision and resolution

The  MR8990  can  measure  minuscule  fluctuations  in  sensor  output  of 

automobiles and voltage fluctuations in batteries, both at high precision 

and resolution. It can accommodate maximum input of 500 V DC. The unit 

is distinguished by its high input resistance. Additionally, the amount of 

space taken up by instruments can be reduced by replacing a bench-style 

DMM with the MR8847A. Systems can be simplified by eliminating the 

need to control multiple instruments. 

Digital Voltmeter Unit MR8990

Measurement range

Effective input 

range

(Guaranteed 

measurement 

accuracy range)

Max.  

resolution

Input  

resistance

Measurement accuracy

NPLC:  

less than 1

NPLC:  

1 or more

5 mV/div

(f.s. = 100 mV)

-120 mV to 120 mV

0.1 μV

100 MΩ

or more

± 0.01% rdg. 

± 0.015% f.s.

± 0.01% rdg. 

± 0.01% f.s.

50 mV/div (f.s. = 1000 mV) -1200 mV to 1200 mV

1 μV

± 0.01% rdg. 

± 0.0025% f.s.

500 mV/div (f.s. = 10 V)

-12 V to 12 V

10 μV

5 V/div

(f.s. = 100 V)

-120 V to 120 V

100 μV

10 MΩ 

± 5%

± 0.025% rdg. 

± 0.0025% f.s.

50 V/div

(f.s. = 1000 V)

-500 V to 500 V

1 mV

 6.5-digit display (resolution: 0.1 μV), 24-bit high resolution

For example, use five 4CH Analog Units to measure 20 analog 

channels and three Logic Units to measure up to 64 channels 
relay on/off signals or PLC (programmable logic controller) 
signals. That’s simultaneous measurement of 84 channels!

Summary of Contents for MR8847A

Page 1: ...hannel voltage and current simultaneously Compatible with multi channel input units Ver 2 00 or later High voltage 1000 V direct input measurement Max 1 MS s high speed sampling 16 bit resolution meas...

Page 2: ...n A D converter Connections between analog input channels and between the input channel and the main unit are isolated by isolation elements So potential differences can be measured without any concer...

Page 3: ...c input channels Add 3 logic input units to record a total of 64 channels at once The waveforms of all channels can be displayed on a single screen making it ideal for timing measurements SSD 128 GB s...

Page 4: ...U8978 can measure multiple channels of output from a variety of sensors The unit is ideal for use in measuring currents of various magnitudes in the development of automobile accessory controls Utiliz...

Page 5: ...even in the confined space of a vehicle s engine compartment Current sensor lineup CT6843 05 CT6877 Output waveform example Sine wave Triangular wave Square wave Pulse wave Ramp down wave Sweep outpu...

Page 6: ...tion Confirmation of UPS switching Transient response measurement of motors Long term stability measurement by recorder function 4 ch current measurement 4 ch voltage measurement 2 ch voltage measurem...

Page 7: ...tput waveforms that you created yourself for testing and measure the results at the same time Record anomalous waveforms Reproduce and output anomalous waveforms Max 15 V output amplifier Generate sim...

Page 8: ...uired data You can use this function to search for anomalies after data has been acquired when it is too difficult to set triggers because it is not possible to predict what types of anomalies might b...

Page 9: ...n a PC connected via LAN You can also use the FTP function to acquire data from the internal memory or from storage media inserted in the MEMORY HiCORDER You can even acquire data from the internal me...

Page 10: ...dead time between measurements A D conversion Store on internal memory Save on devices and media CF card USB memory stick Input Record all sampling data Input SSD Sampling period Recording method Reco...

Page 11: ...for the MR8847 51 64 M words memory capacity Model MR8847 52 256 MW can record four times and Model MR8847 53 512 MW eight times as much On the Continuous setting in recording length total recording t...

Page 12: ...FFT analysis of data measured with the memory function you can use the jog shuttle to specify analysis points while also viewing the calculation results at the same time You can also display both the...

Page 13: ...waveforms that are normally difficult to judge For time axis ranges that are slower than 100 msec div you can even make judgments while loading waveforms This allows you to take the appropriate action...

Page 14: ...100 or 95 of the recording length in 15 stages or in 1 div step settings Numerical calculations Simultaneous calculation for up to 16 selected channels Average value effective rms value peak to peak...

Page 15: ...h 53 min 20 s 2 d 23 h 06 min 40 s 8 h 53 min 20 s 17 h 46 min 40 s 5 d 22 h 13 min 20 s 500 ms div 5 ms 2 h 46 min 40 s 5 h 33 min 20 s 1 d 20 h 26 min 40 s 11 h 6 min 40 s 22 h 13 min 20 s 7 d 09 h...

Page 16: ...ange 100 mV 200 mV 400 mV 1 V 2 V 4 V 10 V 20 V 40 V f s 9 ranges Low pass filter 5 Hz 500 Hz 5 kHz 200 kHz Measurement resolution 1 32 000 of measurement range using 16 bit A D conversion Maximum sam...

Page 17: ...stment range within 10 000 or less Input terminals NDIS connector EPRC07 R9FNDIS via Conversion Cable L9769 and NDIS connector PRC03 12A10 7M10 5 Max rated voltage to ground 30 V rms or 60 V DC with i...

Page 18: ...ensors for use with up to 3 units CURRENT UNIT 8971 2 ch for measuring current using dedicated current sensors 2 CONVERSION CABLES 9318 included for use with up to 4 units TEMP UNIT 8967 2 ch thermoco...

Page 19: ...lution to 200 A range with WAVE RMS output CONNECTION CABLE L9097 output terminal BNC power terminal USB C 1 5 m 4 92 ft length is included AC ADAPTER Z1013 100 V to 240 V AC CONVERSION CABLE CT9901 C...

Page 20: ...0 25 f s 1000 V DC 700 V AC 0 125 mV 4 V f s Yes n a Digital Voltmeter Unit MR8990 Voltage high resolution 2 ch 2 ms n a 24 bit 0 01 rdg 0 0025 f s 500 V DC 0 1 uV 100 mV f s Yes n a 3CH Current Unit...

Reviews: