![Hioki MR8827 Instruction Manual Download Page 409](http://html.mh-extra.com/html/hioki/mr8827/mr8827_instruction-manual_2137533409.webp)
A
4
For Reference
CSV (Text) File Size (Recorder Function)
File size = header size + data size
Calculation method: “Recorder Function” (p. A4)
When saving, CSV (text) files longer than 2 GB are segmented into files of 2 GB each.
Recording
length (div)
Data quantity
Saved channels
1
2
4
8
16
32
100
10,001
450 KB
723 KB
1.2 MB
2.3 MB
4.4 MB
8.7 MB
1,000
100,001
4.4 MB
7.1 MB
12 MB
23 MB
44 MB
87 MB
10,000
1,000,001
44 MB
71 MB
124 MB
231 MB
444 MB
872 MB
50,000
5,000,001
219 MB
353 MB
620 MB
1.1 GB
2.2 GB
4.3 GB
Waveform File Size Calculation Method
MEM File
File size (bytes) = setting size
*1
+ data size
*2
*1: (Setting size) =28672 + 512 × [(number of analog channels) + 4 × (number of logic units) + (number of
waveform calculation channels)]
*2: (Data size) = 2 × [(number of analog channels) + (number of logic units) + 2 × (number of waveform calculation
channels)] × (number of data)
REC File
File size (bytes) = setting size
*1
+ data size
*2
*1: (Setting size) = 28672 + 512 × [(number of analog ch 4 × (number of logic units)]
*2: (Data size) = 4 × [(number of analog channels) + (number of logic units)] × (number of data)
FFT File
File size (bytes) =header size
*1
+ size of time axis data
*2
+ size of mid-term data
*3
*1: (Header size) = 30720 + 512 × [(number of analog channels) + (number of waveform calculation channels) +
(number of FFT calculation channels) + 1]
*2: (Size of time axis data) = [(number of analog channels) + 2 × (number of waveform calculation channels)] ×
(number of data) (*4)
*3: (Size of mid-term data) = [346 × (number of FFT points) + 836] × (number of FFT calculation channels) (*4)
(Depending on the measurement conditions, the file size may also increase or decrease from calculation formula
results)
XYC File
File size (bytes) = header size
*1
+ data size
*2
*1: (Header size) = 24576 + 512 × (number of analog channels)
*2: (Data size) = 2 × (number of saved analog channels) × (number of data)
CSV (Text) File
Memory Function
File size (bytes) = header size
*1
+ data size
*2
*1: (Header size) = 194 + 103 × [(number of saved analog channels) + number of saved logic channels)]
*2: (Data size) = [18 + 14 × (number of saved analog channels) + 2 × (number of logic channels)] × (number of
data)
Recorder Function
File size (bytes) = header size
*1
+ data size
*2
*1: (Header size) = 194 + 130 × [(number of saved analog channels) + (number of saved logic channels)]
*2: (Data size) = [18 + 28 × (number of saved analog channels) + 4 × (number of logic channels)] × (number of
data)
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......