278
FFT Analysis Modes
Transfer Function
From the input and output signals, the transfer function (frequency characteristic) of a
measurement system can be obtained. It can also be displayed as a Nyquist plot.
Main uses:
•
To inspect a filter’s frequency characteristic
• To inspect the stability of a feedback control system (using the Nyquist plot)
• To inspect the resonance characteristic of an object using an impulse hammer and pick-up sensor
About the Functions, refer to“13.9.2 Analysis Mode Functions” (p. 289) or Refer to “Linear Time-Invariant Systems”
Axis
Display type
Description
X axis
Linear
Displays the frequency on a linear scale
Log
Displays the frequency on a logarithm scale
Nyquist display
Displays the real-number component of the input-output ratio.
Y axis
Lin-Mag
Displays the input-output ratio linearly (dimensionless units).
Log-Mag(logarithm) Displays the input-output ratio as dB values.
Lin-Real
Displays the real-number component of the input-output ratio. (dimensionless
units)
Lin-Imag
Displays the imaginary component of the input-output ratio. (dimensionless
units)
Nyquist display
Displays the imaginary component of the input-output ratio.
Waveform Example
Normal display
X axis: Log
Y axis: Lin-Mag
Normal display
X axis: Log
Y axis: Log-Mag
Normal display
X axis: Log
Y axis: Lin-Real
Normal display
X axis: Log
Y axis: Lin-Imag
Nyquist display
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......