252
FFT Analysis Condition Settings
13.3.7
Average Analysis Results (Waveform Averaging)
The averaging function calculates the average of the values obtained from multiple measurements of a
periodic waveform.
This can reduce noise and other non-periodic signal components. Averaging can be applied to a time-
domain waveform or to a spectrum.
Procedure
Press the
STATUS
key to open the Status screen, and then select the
[Status]
sheet.
1
Enable averaging.
Move the cursor to
[Average]
.
Off
Averaging is disabled.
(Default setting)
Linear (Time
axis)
Time-domain waveforms are summed and
averaged and then calculation is performed.
Expo. (Time
axis)
The exponential mean of time-domain
waveforms is determined and then calculation
is performed.
Linear
(Frequency)
Frequency-domain waveforms are summed
and averaged and then the calculation result
is output.
Expo.
(Frequency)
The exponential mean of frequency-domain
waveforms is determined and then the
calculation result is output.
Peak
(Frequency)
The maximum value of frequency-domain
waveforms is retained.
1
2
About averaging calculation
formulas
Refer to “Averaging” (p. A19).
When averaging and auto-saving
or auto-printing are enabled at the
same time
Data is saved or printed after the
specified number of values has been
averaged.
Refer to “Trigger Modes and Averaging”
2
Select the count for averaging.
Move the cursor to
[Number]
.
Select the number of measurements to be averaged.
Setting range: 2 to 10,000
• After measuring with averaging enabled, display is not available when the channel is changed. Also,
when the analysis mode is changed, the analysis modes that can be displayed are limited.
• When averaging is performed with analysis mode disabled (Off), no trace will be displayed when the
analysis mode is changed after measurement.
• When
[Format]
is set to
[Running spectrum]
,
[Average]
cannot be used.
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......