254
FFT Analysis Condition Settings
■
Trigger Modes and Averaging
If the trigger mode is
[Single]
or the calculation setting is
[Once]
Measurement continues until the specified number of averaging points is acquired
When the acquired count < the specified count
*1
When the acquired
count = the specified count
Start Measurement
START
key
End
*2
Trigger
criteria met
Input
Signal
Trigger
Wait
Waveform acquisition
(specified no. of points)
Averaging
FFT
analysis
FFT
analysis
(Spectrum averaging) (Time-domain waveform averaging)
Display
*3
*1. Awaiting trigger continues until the specified count is reached.
*2. Measurement stops automatically when the specified count is reached. If measurement was interrupted by
the
STOP
key, the averaging result up to that point is displayed.
*3. If
[Reference]
is
[New Data]
and automatic saving or automatic print is
[On]
, data is saved or printed
when the specified number of times is reached.
If the trigger mode is
[Continue]
or the calculation setting is
[Repeat]
Measurement continues after the specified averaging count has been acquired.
When the specified averaging count is exceeded, averaging is repeated and measurement
continues until the
STOP
key is pressed.
Repeats until the
STOP
key is pressed
End
*1
Start Measurement
START
key
Trigger
criteria met
Input
Signal
Awaiting
Trigger
Averaging
Display
*2
FFT
analysis
FFT
analysis
(Spectrum averaging) (Time-domain waveform averaging)
STOP
key
Waveform acquisition
(specified no. of points)
*1. When stopped before the specified count, the average up to that point is displayed.
*2. If
[Reference]
is
[New Data]
and automatic saving or automatic print is
[On]
, data is saved or printed
when the specified number of times is reached.
When the trigger mode is
[Auto]
• For time-domain waveforms
When the
START
key is pressed, after a certain interval data will be acquired even if trigger
criteria are not met. This means that if averaging is applied to an asynchronous signal, the
resulting data is meaningless.
• For time-domain waveforms
When the
START
key is pressed, measurement starts. Even if the trigger criteria are not met, the
specified amount of data is acquired, and after FFT analysis, the results are averaged.
When the specified averaging count is exceeded, averaging is repeated and measurement
continues until the
STOP
key is pressed.
Repeats until the
STOP
key is pressed
End
*
Start Measurement
START
key
After a certain interval,
acquisition starts even if
trigger criteria are not met
Input
Signal
Awaiting
Trigger
Averaging
Display
FFT
analysis
FFT
analysis
(Spectrum averaging) (Time-domain waveform averaging)
STOP
key
Waveform acquisition
(specified no. of points)
* When stopped before the specified count, the average up to that point is displayed.
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......