Configuring External Input/Output (I/O) Terminals
Unit
Measure
Channel
Trigger
Alarm
Calculation
System
External trigger input
This section describes how to input a signal from an external source to activate a trigger.
You can use a signal from another device to start measurement with the instrument. The external
input I/O 3 terminal is used for this functionality.
>
>
1
1
2
2
1
Under
[External input 3]
, set the terminal functionality to
[Trigger input]
.
When
[External trigger]
is set to
[ON]
, the parameter will be set to
[Trigger input].
See “2.5 Applying Triggers Based on External Sources” (p. 94).
2
Select the edge at which to activate the trigger.
↑
Activates the trigger at the rising edge when changing from low level to high level.
↓
Activates the trigger at the falling edge when changing from high level to low level.
The trigger will also activate if the I/O 3 and GND terminals are shorted.
It is recommended to actually input the signal and verify whether the external trigger operates.
8
External Control (EXT
. I/O)