![Hioki LR8432-20 Instruction Manual Download Page 57](http://html1.mh-extra.com/html/hioki/lr8432-20/lr8432-20_instruction-manual_2137139057.webp)
3.3 Setting Configuration 1 – Recording Settings
49
3
3
C
h
apt
er 3
Set
ting
be
fo
re M
easurement
Displayed waveforms can be expanded or compressed
along the horizontal axis. Shorter setting values corre-
spond to greater magnification. This setting can be
changed both during and after measurement without
affecting acquired data (p. 99).
Setting the Display Time Base (horizontal axis magnification, set
as needed)
1 div
Select from the
[Time/DIV]
setting options.
Setting options:
100ms
∗
, 200ms, 500ms, 1s, 2s, 5s, 10s, 20s, 30s,
1min, 2min, 5min, 10min, 20min, 30min, 1h, 2h, 5h,
10h, 12h, 1day
[/DIV]
(
∗
default setting)
When auto save is enabled for waveforms, 100 ms, 200 ms
and 500 ms settings are automatically changed to 1 s when
measurement starts, and the shorter values cannot be se-
lected while measuring.