5.1 Triggering Measurement Start and Stop
105
3
5
Chapter 5 S
p
ecif
y
ing Crite
ria
for M
easurement
s
Using a Window Trigger
An input signal level “window” within which recording will (or will
not) occur can be defined by upper and lower threshold levels.
You can select whether measurement starts or stops when the
input signal level enters (In) or exits (Out) of this window.
Upper and lower thresholds can be verified in the monitor or on
the Waveform screen. When scaling is enabled, the scaled
threshold values are displayed.
Make trigger settings on the
[CH]
screen of the displayed chan-
nel.
See:
"Key Setting Procedure" (p .103)
T
T
T
T
IN
OUT
1
Select
[Window]
as the start/stop trigger type.
2
Select trigger timing (window type).
3
Enter the upper/lower threshold value.
Setting options: (
∗
: default setting)
In
∗
Trigger when the input signal level enters the
window defined by the upper/lower thresholds.
Out
Trigger when the input signal level exits the win-
dow defined by the upper/lower thresholds.
1
2
3
3
Scaled values are displayed
when scaling is enabled.
Summary of Contents for LR8400-20
Page 1: ......
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 14...
Page 96: ...3 9 Viewing and Editing with the All Channel Settings List 84...
Page 112: ...4 3 Marking Waveforms and Searching Marks Search Function 100...
Page 156: ...6 5 Transferring Data to a PC USB Drive Mode 144...
Page 238: ...10 8 About Communications Commands 226...
Page 256: ...11 3 LR8501 Universal Unit Specifications 244...
Page 262: ...12 3 Disposing of the HiLogger 250...
Page 296: ...Index Index 4...
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......