5.1 Triggering Measurement Start and Stop
102
There are three ways to set the criteria to start and stop recording according to
waveform slope.
Select whether to trigger at the rising or falling edge of the input signal.
5.1
Triggering Measurement Start and Stop
Types of Trigger Criteria
Type of Analog
Trigger
Trigger Example
Description
Level Trigger
[Level]
Recording starts (or stops) when the
input signal crosses the specified trig-
ger level (voltage).
Window Trigger
[In]
Recording starts (or stops) when the
input signal enters a range defined by
upper and lower thresholds.
[Out]
Recording starts (or stops) when the
input signal exits a range defined by
upper and lower thresholds.
Logic Trigger
[1]
Measuring starts and stops when the
input signal logic switches from Low to
High.
[0]
Measuring starts and stops when the
input signal logic switches from High to
Low.
[X]
Signal is ignored. No triggering occurs.
Trigger Level
Input
Trigger Slope
[
↓
]
[
↑
]
T
T
Upper
Threshold
Lower
Threshold
T
T
Upper
Threshold
Lower
Threshold
T
T
T T
T
High
Low
T T
High
Low
High
Low
Summary of Contents for LR8400-20
Page 1: ......
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 14...
Page 96: ...3 9 Viewing and Editing with the All Channel Settings List 84...
Page 112: ...4 3 Marking Waveforms and Searching Marks Search Function 100...
Page 156: ...6 5 Transferring Data to a PC USB Drive Mode 144...
Page 238: ...10 8 About Communications Commands 226...
Page 256: ...11 3 LR8501 Universal Unit Specifications 244...
Page 262: ...12 3 Disposing of the HiLogger 250...
Page 296: ...Index Index 4...
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......