7.2 Measurement Specifications
1
2
3
4
5
6
7
8
9
10
11
付録
索引
147
Values measured: Insulation resistance, leakage current,
voltage, and temperature
7.2.1 Insulation Resistance Measurement
7.2 Measurement Specifications
Measurement
test voltage
Selectable range: 250 V DC to 5.00 kV
DC
Setting method:
• Choose from test voltage presets (250 V,
500 V, 1 kV, 2.5 kV, 5 kV)
• Fine adjustment (between 250 V and 1
kV with a resolution of 25 V or between 1
kV and 5 kV with a resolution of 100 V.)
Output voltage
accuracy
• -0% and +10% of setting
• Applies when the instrument measures
a resistance equal to or higher than the
result of division of test voltage (set
value) by rated measuring current.
Short-circuit
current
2 mA or less
Output voltage
monitor
function
Display range: 0 V to 999 V, 0.98 kV to
5.50 kV
Monitored value accuracy:
±
5% rdg.
±
5 dgt.
(Actual output voltage is within the tolerance
of the output voltage accuracy given above.)
Measuring
range
Resistance obtained by dividing the value
of the range from 0.00 M
to the test volt-
age (set value) by 0.5 nA (Measuring
range varies according to test voltage.)
*Rated measuring current:
Electric current that can be generated with the set
test voltage is maintained.
Test voltage
(setting)
Rated measuring
current*
(Tolerance: -0%,+10%)
250 V to 1.00 kV
1 mA
1.10 kV to 2.50 kV
0. 5 mA
2.60 kV to 5.00 kV
0.25 mA
Summary of Contents for IR3455
Page 2: ......
Page 24: ...Operating Precautions 18...
Page 110: ...4 4 Step Voltage Test 104...
Page 160: ...7 3 9750 01 02 03 11 12 13 and 9751 01 02 03 ALLI GATOR CLIPs Specifications 154...
Page 180: ...Appendix 4 Temperature Compensation Table 174...
Page 181: ...1 2 3 4 5 6 7 8 9 10 11...
Page 182: ......
Page 183: ......
Page 184: ......