7.2 Short Circuit Compensation
285
7
Chap
ter 7 Erro
r Com
pensatio
n
Acquire the compensation values at the set measurement frequencies. Measurement frequencies can be set
for up to five points.
7.2.2
Spot Compensation
Procedure
1
Press
to confirm the frequency for compensation.
Select
in the short circuit compen-
sation screen, and press
to confirm
the selection.
3
4
Short Compensation Setting
Spot Compensation Value Setting
2
Spot Compensation
Until one of these keys is pressed for input of a nu-
merical value, the previous frequency for which SPOT
compensation was performed is displayed.
Use
or
to select the compensation
point you want to set or edit, and then press
.
• Settable range: DC, 4 Hz to 5.000 MHz
• If a value in excess of 5 MHz is entered, the
measurement frequency will automatically set to
5 MHz.
• If a value less than 4 Hz is entered, the measure-
ment frequency will automatically set to 4 Hz.
• If you make a mistake during input:
press
to cancel the input and start again.
Use the numeric keypad to enter a frequency
for compensation.
When you do not want to perform compen-
sation:
Press .
Compensation is not performed and the compensa-
tion screen is redisplayed.
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......