5.10 Equivalent Circuit Analysis Function
248
4
Oscillation Mode Settings
Select the oscillation mode.
Electromechanical coupling coefficient for pla-
nar oscillation
Electromechanical coupling coefficient for
long-side extension oscillation
Electromechanical coupling coefficient for lon-
gitudinal oscillation
Electromechanical coupling coefficient for
thickness-mode oscillation
Electromechanical coupling coefficient for
shearing oscillation
Kr
f
p
f
s
–
a f
s
b
f
p
f
s
–
+
--------------------------------------------------------
=
K31
2
---
f
p
fs
-----
2
---
f
p
fs
-----
2
---
f
p
fs
-----
tan
–
--------------------------------------------------------
=
K33
2
---
fs
f
p
-----
2
---
fs
f
p
-----
cot
=
Kt
2
---
fs
f
p
-----
2
---
fs
f
p
-----
cot
=
K15
2
---
fs
f
p
-----
2
---
fs
f
p
-----
cot
=
5
Select the frequency type
Frequency Type Setting
Select the resonant frequency to use when calculating
the electromechanical coupling coefficient.
Selects the series/parallel resonant frequency.
Selects the resonant/anti-resonant frequency.
(Substitutes
f
r
for
f
s
and
f
a
for
f
p
in the
equations in Step
4
.)
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......