Hioki IM3523 Instruction Manual Download Page 7

1.1  RS-232C Specifications

3

When using hardware (RTS/CTS control):

When the data in the receive buffer exceeds 85% of
the buffer, CA(RTS) is set to OFF and the controller is
notified that there is not much space remaining in the
buffer. 

Processing of the data in the buffer continues, and
then CA(RTS) is set to ON and the controller is noti-
fied that there is sufficient remaining space in the buf-
fer when the amount of data becomes less than 25%.

When using software (XON/XOFF control):

When the data in the receive buffer exceeds 75% of
the buffer, XOFF(13H) is sent and the controller is notified that there is not much space remaining in
the buffer. 

Processing of the data in the buffer continues, and then XON(11H) is sent and the controller is notified
that there is sufficient remaining space in the buffer when the amount of data becomes less than 25%. 

When using hardware (RTS/CTS control):

When CB(CTS) is confirmed to be OFF, the sending of data is halted. When it is confirmed to be ON,
the sending of data is resumed. 

When using software (XON/XOFF control): 

When XOFF is received, the sending of data is halted. When XON is received, the sending of data is
resumed. 

Specifications

Chapter 1

1.1 RS-232C Specifications

Transmission Method

 

Communication method: Full duplex
Synchronous method: Start-stop synchronization

Transmission Speed

9600 bps, 19200 bps, 38400 bps, 57600 bps

Data Bits

 8 bits

Parity

None

Stop bit

1 bits

Message terminator 
(delimiter)

CR+LF, CR

Flow control

Hardware (RTS/CTS control), software (XON/XOFF control)
“Handshake (About Buffer Flow Control)” (p. 3)
IM7580: Software (XON/XOFF control only)

Electrical 
Specifications

Input voltage 
level
Output voltage 
level

 5 to 15 V ........ ON
-15 to -5 V ....... OFF
 5 to 9 V .......... ON
-9 to -5 V  ........ OFF

Handshake (About Buffer Flow Control)

Control during Receiving 

CA(RTS)

OFF

ON

25%

85%

Input Buffer 
Usage 
Amount

Buffer Space

75%

Send XOFF

Send XON

Control during Sending

HIOKI IM3570A983-08

Summary of Contents for IM3523

Page 1: ...nstruction Manual EN Dec 2022 Revised edition 8 IM3570A983 08 22 12H IM3523 IM3533 LCR METER IM3533 01 IM3536 IMPEDANCE ANALYZER IM3570 IM7580 CHEMICAL IMPEDANCE ANALYZER IM3590 IM3523A 600329708 HIOK...

Page 2: ...HIOKI IM3570A983 08...

Page 3: ...g the USB Cable 24 3 3 GP IB Connection and Settings IM3523 only when connected to the Z3000 25 Connecting the GP IB Cable 25 Setting GP IB 25 3 4 RS 232C Connection and Settings IM3523 only when conn...

Page 4: ...and Connection 55 LAN Settings 55 Connecting a LAN Cable 60 5 4 GP IB Connection and Settings when connected to the Z3000 61 Connecting the GP IB Cable 61 Setting GP IB 61 5 5 RS 232C Connection and S...

Page 5: ...site https www hioki com global support download Product registration Register your product in order to receive important product information https www hioki com global support myhioki registration Sa...

Page 6: ...Press and quickly release the right button of the mouse Double click Quickly click the left button of the mouse twice Drag While holding down the left button of the mouse move the mouse and then rele...

Page 7: ...When using hardware RTS CTS control When CB CTS is confirmed to be OFF the sending of data is halted When it is confirmed to be ON the sending of data is resumed When using software XON XOFF control...

Page 8: ...ons RL1 Supports all remote local functions PP0 Parallel poll functions are not supported DC1 Supports all device clear functions DT1 Supports all device trigger functions C0 Controller functions are...

Page 9: ...isconnecting an interface connector Otherwise an electric shock accident may occur To avoid damage to the instrument do not short circuit the terminal and do not input voltage to the terminal Failure...

Page 10: ...nterface the icon for the set interface is displayed on the right side of the screen Screen Displayed while Setting Interfaces When RS 232C is set When GP IB is set When USB is set When LAN is set Whe...

Page 11: ...compatible with the connectors on both the instrument and the controller The I O connector is a DTE Data Terminal Equipment configuration 6 7 8 9 1 2 3 4 5 Connector D sub Pin No Interchange Circuit N...

Page 12: ...the setting from any of mode and mode only IM3570 Press 2 RS 232C Settings 3 Press to confirm the setting 4 RS 232C Settings Select the terminator setting CR LF CR Select the baud rate setting Select...

Page 13: ...nector 2 3 GP IB Connection and Settings Connecting the GP IB Cable Setting GP IB Recommended cable 9151 02 GP IB connection cable 2 m Procedure LCR Measurement Screen 1 Interface Settings You can con...

Page 14: ...IB Connection and Settings 10 Press 2 GPIB Setting 3 GPIB Setting Use or to set the GP IB address Select the terminator setting LF with EOI LF with CR EOI Press to confirm the setting 4 HIOKI IM3570A...

Page 15: ...ed CD or our web site http www hioki com The USB driver is compatible with the Windows 7 32 bit 64 bit version Windows 8 32 bit 64 bit version Windows 10 32 bit 64 bit version and Windows 11 64 bit ve...

Page 16: ...ly available USB cable to the USB port of the instrument Connecting the USB Cable 3 USB Setting Select the terminator setting CR LF CR Press to confirm the setting 4 USB cable commercially available p...

Page 17: ...erent IP address than the computer Subnet Mask 255 255 255 0 Default Gateway OFF 0 0 0 0 2 5 LAN Settings and Connection LAN Settings Make these settings before connecting to a network Changing settin...

Page 18: ...and Connection 14 Procedure LCR Measurement Screen 1 Interface Settings You can configure the setting from any of mode and mode only IM3570 Press 2 LAN Settings LAN Settings 3 Select the IP address HI...

Page 19: ...0 255 255 224 000 255 255 255 224 240 000 000 000 255 240 000 000 255 255 240 000 255 255 255 240 248 000 000 000 255 248 000 000 255 255 248 000 255 255 255 248 252 000 000 000 255 252 000 000 255 25...

Page 20: ...ay does not need to be set for example when connecting the instrument and computer on a one to one basis using a cross ca ble leave this set to OFF Default gateway Settings 8 Use or to set the default...

Page 21: ...er Settings 10 Use or to set the port number to use for communication commands Press to confirm the setting Settable range 1024 to 65535 LAN Settings 11 Select the terminator setting CR LF CR Press to...

Page 22: ...ecting a LAN Cable When connecting the instrument to your LAN using a LAN cable of more than 30 m or with the cable laid outdoors take appropriate countermeasures that include install ing a surge prot...

Page 23: ...ode remote operation state and the keys on the LCD are disabled 2 6 Remote Mode Canceling Remote Mode All of the keys except are disabled Remote Mode State Procedure 2 Press to return to the normal st...

Page 24: ...2 6 Remote Mode 20 HIOKI IM3570A983 08...

Page 25: ...ircuit the terminal and do not input voltage to the terminal Failure to fasten the connectors properly may result is sub specification perfor mance or damage to the equipment To avoid damage do not di...

Page 26: ...USB and LAN inter faces There are the following two communication methods To enable communication the communication condi tions need to be set on the instrument The instrument is communication class...

Page 27: ...m IM3523A When the instrument is connected to a computer the USB driver is automatically installed Since the OS standard driver is installed it is not necessary to install another driver The USB drive...

Page 28: ...o the USB port of the instrument 3 Select the terminator setting 4 Connecting the USB Cable Select 1 Select the terminator setting 2 You will return to the measure ment screen USB cable commercially a...

Page 29: ...GP IB connector 3 3 GP IB Connection and Settings IM3523 only when connected to the Z3000 Connecting the GP IB Cable Setting GP IB 1 Open the SYSTEM screen 2 Select GP IB as the interface Recommended...

Page 30: ...nected to the Z3000 26 3 Select the terminator setting 4 Set the GP IB address Valid setting range 0 to 30 5 Select 1 LF with CR EOI 2 LF with EOI Select 1 Select the GP IB address setting 2 You will...

Page 31: ...troller DTE use a crossover cable compatible with the connectors on both the instrument and the controller The I O connector is a DTE Data Terminal Equipment configuration 6 7 8 9 1 2 3 4 5 Connector...

Page 32: ...the SYSTEM screen 2 Select RS 232C as the interface 3 Select from the following baud rate setting 9600 19200 38400 57600 4 Select the terminator setting Open the SYSTEM screen Select 1 Select RS 232C...

Page 33: ...d to the Z3001 29 5 Select the handshake setting 6 Select 1 Select the handshake setting 2 OFF No flow control HARD Hardware RTS CTS control XON OFF Software XON XOFF control BOTH Hardware software Yo...

Page 34: ...erent IP address than the computer Subnet Mask 255 255 255 0 Default Gateway OFF 0 0 0 0 3 5 LAN Settings and Connection IM3523 needs to be connected to the Z3002 LAN Settings Make these settings befo...

Page 35: ...255 192 000 000 255 255 192 000 255 255 255 192 224 000 000 000 255 224 000 000 255 255 224 000 255 255 255 224 240 000 000 000 255 240 000 000 255 255 240 000 255 255 255 240 248 000 000 000 255 248...

Page 36: ...he default gateway 2 Set the default gateway to OFF 000 000 000 000 If the default gateway does not need to be set for example when connecting the instrument and computer on a one to one basis using a...

Page 37: ...ument to a single computer connect the instrument to the computer Connecting a LAN Cable On Performing communication with 100BASE Off Performing communication with 10BASE Flashes when data is being ex...

Page 38: ...n the mode becomes remote mode remote operation state and the keys on the LCD are disabled 3 6 Remote Mode Remote status Canceling Remote Mode 1 2 Keys other than F1 are disabled You will return to th...

Page 39: ...ument do not short circuit the terminal and do not input voltage to the terminal Failure to fasten the connectors properly may result is sub specification perfor mance or damage to the equipment To av...

Page 40: ...et an interface the icon for the set interface is displayed on the right side of the screen Screen Displayed while Setting Interfaces When USB is set When GP IB is set When RS 232C is set When LAN is...

Page 41: ...ur web site http www hioki com The USB driver is compatible with the Windows 7 32 bit 64 bit version Windows 8 32 bit 64 bit version Windows 10 32 bit 64 bit version and Windows 11 64 bit version oper...

Page 42: ...ly available USB cable to the USB port of the instrument Connecting the USB Cable 3 USB Setting Select the terminator setting CR LF CR Press to confirm the setting 4 USB cable commercially available p...

Page 43: ...connector 4 3 GP IB Connection and Settings when connected to the Z3000 Connecting the GP IB Cable Setting GP IB Recommended cable 9151 02 GP IB connection cable 2 m You can configure the setting fro...

Page 44: ...and Settings when connected to the Z3000 40 Press 2 GPIB Setting 3 GPIB Setting Use or to set the GP IB address Select the terminator setting LF with EOI LF with CR EOI Press to confirm the setting 4...

Page 45: ...use a crossover cable compatible with the connectors on both the instrument and the controller The I O connector is a DTE Data Terminal Equipment configuration 6 7 8 9 1 2 3 4 5 Connector D sub Pin No...

Page 46: ...You can configure the setting from any of mode mode and mode Press 2 RS 232C Settings 3 Press to confirm the setting 4 RS 232C Settings Select the handshake setting No flow control Hardware RTS CTS co...

Page 47: ...erent IP address than the computer Subnet Mask 255 255 255 0 Default Gateway OFF 0 0 0 0 4 5 LAN Settings and Connection when connected to the Z3002 LAN Settings Make these settings before connecting...

Page 48: ...ion when connected to the Z3002 44 You can configure the setting from any of mode mode and mode Procedure LCR Measurement Screen 1 Interface Settings Press 2 LAN Settings LAN Settings 3 Select the IP...

Page 49: ...255 224 000 000 255 255 224 000 255 255 255 224 240 000 000 000 255 240 000 000 255 255 240 000 255 255 255 240 248 000 000 000 255 248 000 000 255 255 248 000 255 255 255 248 252 000 000 000 255 252...

Page 50: ...default gateway does not need to be set for example when connecting the instrument and computer on a one to one basis using a cross ca ble leave this set to OFF Default gateway Settings 8 Use or to s...

Page 51: ...2 47 Port number Settings 10 Use or to set the port number to use for communication commands Press to confirm the setting Settable range 1024 to 65535 LAN Settings 11 Select the terminator setting CR...

Page 52: ...with cross over adapter up to 100 m Hioki 9642 LAN Cable option When connecting the instrument to a single computer connect the instrument to the computer Connecting a LAN Cable On Performing communi...

Page 53: ...ode remote operation state and the keys on the LCD are disabled 4 6 Remote Mode Canceling Remote Mode All of the keys except are disabled Remote Mode State Procedure 2 Press to return to the normal st...

Page 54: ...4 6 Remote Mode 50 HIOKI IM3570A983 08...

Page 55: ...erminal and do not input voltage to the terminal Failure to fasten the connectors properly may result is sub specification perfor mance or damage to the equipment To avoid damage do not disconnect the...

Page 56: ...When you set an interface the icon for the set interface is displayed on the right side of the screen Screen Displayed while Setting Interfaces When USB is set When GP IB is set When RS 232C is set Wh...

Page 57: ...an be downloaded from the bundled CD or our web site http www hioki com The USB driver is compatible with the Windows 7 32 bit 64 bit version Windows 8 32 bit 64 bit version Windows 10 32 bit 64 bit v...

Page 58: ...ly available USB cable to the USB port of the instrument Connecting the USB Cable 3 USB Setting Select the terminator setting CR LF CR Press to confirm the setting 4 USB cable commercially available p...

Page 59: ...erent IP address than the computer Subnet Mask 255 255 255 0 Default Gateway OFF 0 0 0 0 5 3 LAN Settings and Connection LAN Settings Make these settings before connecting to a network Changing settin...

Page 60: ...tings and Connection 56 You can configure the setting from any of mode and mode Procedure LCR Measurement Screen 1 Interface Settings Press 2 LAN Settings LAN Settings 3 Select the IP address HIOKI IM...

Page 61: ...0 255 255 224 000 255 255 255 224 240 000 000 000 255 240 000 000 255 255 240 000 255 255 255 240 248 000 000 000 255 248 000 000 255 255 248 000 255 255 255 248 252 000 000 000 255 252 000 000 255 25...

Page 62: ...way does not need to be set for example when connecting the instrument and computer on a one to one basis using a cross ca ble leave this set to OFF Default gateway Settings 8 Use or to set the defaul...

Page 63: ...er Settings 10 Use or to set the port number to use for communication commands Press to confirm the setting Settable range 1024 to 65535 LAN Settings 11 Select the terminator setting CR LF CR Press to...

Page 64: ...e with cross over adapter up to 100 m Hioki 9642 LAN Cable option When connecting the instrument to a single computer connect the instrument to the computer Connecting a LAN Cable On Link up Off Link...

Page 65: ...IB connector 5 4 GP IB Connection and Settings when connected to the Z3000 Connecting the GP IB Cable Setting GP IB Recommended cable 9151 02 GP IB connection cable 2 m You can configure the setting f...

Page 66: ...and Settings when connected to the Z3000 62 Press 2 GPIB Setting 3 GPIB Setting Use or to set the GP IB address Select the terminator setting LF with EOI LF with CR EOI Press to confirm the setting 4...

Page 67: ...sub 4 40 attaching screws To connect the instrument to a controller DTE use a crossover cable compatible with the connectors on both the instrument and the controller The I O connector is a DTE Data...

Page 68: ...en 1 Interface Settings You can configure the setting from any of mode and mode Press 2 RS 232C Settings 3 Press to confirm the setting 4 RS 232C Settings Select the handshake setting No flow control...

Page 69: ...ode remote operation state and the keys on the LCD are disabled 5 6 Remote Mode Canceling Remote Mode All of the keys except are disabled Remote Mode State Procedure 2 Press to return to the normal st...

Page 70: ...5 6 Remote Mode 66 HIOKI IM3570A983 08...

Page 71: ...irtual COM Port Device Manager starts The procedure to start Device Manager differs depending on the version of the Windows operat ing system For details refer to Help of the operating system Check Ch...

Page 72: ...Appendix 1 Checking the USB Virtual COM Port A2 HIOKI IM3570A983 08...

Page 73: ...HIOKI IM3570A983 08...

Page 74: ...HIOKI IM3570A983 08...

Page 75: ...HIOKI IM3570A983 08...

Page 76: ...www hioki com HIOKI IM3570A983 08...

Reviews: