12
―――――――――――――――――――――――――――
Chapter 2 Making Measurements
――――――――――――――――――――――――
NOTE
2.3 Continuous Measurement
2.4 How to Specify the Thermal
Emissivity(
ε
)
1. With the power source off, the product is put
into the continuous measurement mode by
holding the
MODE
key down and then
pressing the
MEAS
key.
2. Anytime
MEAS
key is pressed,
MEAS
and
HOLD
modes toggled.
・
An auto power off function does not operate
in the continuous measurement mode. Be
sure to return to
HOLD
mode. The power goes
off after 15 seconds.
・
During the continuous measurement mode,
the laser marker (3415-02) or LED marker
(3416-02) is not turned on. (Marker is turned
on whenever
MEAS
key is pressed.)
1. Stick black body tape on the object whose
temperature is to be measured, or spray it
with black body spray.
2. Set the thermal emissivity value(
ε
) on the
display screen to the value indicated on black
body tape (black body spray).
3. Press the
MEAS
key, so as to measure the
temperature (T
real
) of the part on which black
body tape (or black body spray) is applied.