29
Inspection before Use
2.4 Inspection before Use
Before using the instrument, verify that it operates normally to ensure
that no damage occurred during storage or shipping. If you find any
damage, contact your authorized Hioki distributor or reseller.
Check items
Solution
Is the battery level sufficient?
Check the battery level indicator on the upper
right corner while the power has been turned
on. If
is blinking, the battery level is low.
Replace the batteries with new ones. (p.25)
Is there any missing segment
in the display items?
Check by activating the LCD all-on display.
If there is a missing segment, the device
needs to be repaired.
Is there any damage or
crack in the device?
Conduct visual checking.
If damage is found, the device should not
be used and needs to be repaired as it can
cause electric shock.
Is there any foreign material
(sand etc.) inside the
measurement terminals?
Remove all foreign materials if any. If it cannot
be removed, the device needs to be repaired.
Is there any damage or
exposed internal white part
or metal in the coating of
measurement cable or test
lead?
If damage is found, it should not be used and
needs to be replaced as it can cause electric
shock.