16
Menu settings
In the measurement screen, press and hold the A/C/CAL key
to switch to the menu screen (not available during processing).
Range setting
Press the START/STOP key or the MODE key to select the fol-
lowing range settings.
Wide:
Measurement range: 30 to 137 dB,
L
Cpeak
cannot
be processed
Peak:
Measurement range: 65 to 137 dB,
L
Cpeak
can be
processed
Even if the frequency weighting characteristic is
A,
L
Cpeak
is processed as C.
When wide range is
selected
Flashes every
0.5 seconds
Press the A/C/CAL key to switch to the time weighting char-
acteristic settings.
Press and hold the A/C/CAL key to return to the measurement
screen.
When peak range is
selected
Flashes every
0.5 seconds
www.
GlobalTestSupply
.com
Find Quality Products Online at: