Hioki CT7631 Instruction Manual Download Page 4

5

6

7

8

CT7631

CT7636

CT7642

Jaw  
dimensions

Approx. 66W × 

13D mm (2.60″W 

× 0.51″D)

Approx. 69W × 

14D mm (2.72″W 

× 0.55″D)

Approx. 92W × 

18D mm (3.62″W 

× 0.71″D)

Mass

Approx. 250 g
(

8.8

 oz.)

Approx. 320 g
(11.3 oz.)

Approx. 510 g
(

18.0

 oz.)

Cable length

Approx. 2.5 m (98.43″

)

Product  
warranty period

3 years

Accessories

Instruction Manual, 
Current Sensor Operating Precautions

Output Specifications and Measurement 

Specifications

(1)

Basic specifications

CT7631

CT7636

CT7642

Output 
connector

Hioki PL14

Rated  
measurement 
current

100 A AC/DC

600 A AC/DC

2000 A AC/DC

Output rate

1 mV/ A

1 mV/ A

0.1 mV/ A

Maximum 
measurement 
current

RMS value, continuous:

see “Frequency deratings” below.

Peak value 

(

under the RMS value conditions described above.):

150 A peak

900 A peak

2840 A peak

Frequency 
band

DC to 10 kHz (

3 dB)

Measurable 
conductor 
diameter

φ

33 mm (

φ

1.30″)

or less

φ

33 mm (

φ

1.30″)

or less

φ

55 mm (

φ

2.17″)

or less

Maximum rated 
voltage to earth

600 V AC/DC
(Measurement 
category IV)

1000 V AC/DC 
(Measurement category III)
600 V AC/DC 
(Measurement category IV)

Anticipated transient overvoltage: 8000 V

CT7631 Frequency derating

Measurement current [A]

Frequency [Hz]

CT7636 Frequency derating

Measurement current [A]

Frequency [Hz]

CT7642 Frequency derating

Measurement current [A]

Frequency [Hz]

(2)

Accuracy specifications

f.s.:  The rated measurement current.
rdg.: The value currently being measured and indicated on the measuring

instrument.

CT7631

CT7636

CT7642

Conditions of 
guaranteed  
accuracy

Guaranteed accuracy period:

3 years

Guaranteed accuracy period 
after adjustment made by Hioki:

3 years

Opening and closing of the jaw:

30000 times or less

Accuracy guarantee for  
temperature and humidity:

23°C±5°C (73°F±9°F), 

80% RH or less

After performing zero-adjustment with the instrument to 
which the device is connected

Accuracy of AC measurement guaranteed for sine wave 
inputs

Measurement 
accuracy

See separate table.

Temperature 

coefficient

In the operating temperature range, add 0.1 × specified 

accuracy/°C (at temperatures other than 23°C±5°C).

Effect of radiated radio-frequency electromagnetic field

15% f.s. at 10 V/m

Effect of conducted radio-frequency electromagnetic field

10% f.s.

 at 3 V

Effect of conductor position (deviation from center)

Within ±1.5%

Within ±2.0%

Within ±1.0%

Effect of external magnetic field (400 A/m, DC)

Within ±1.5% f.s. Within ±0.5% f.s. Within ±0.2% f.s.

Maximum cord 
extension 
length

100 m (Depends on the instrument to which the device is 
to be connected.)

CT7631 Measurement accuracy

Frequency

DC

45 

 f 

≤ 

66

(Hz)

DC < f <

 

45,

66 < f 

≤ 

500

(Hz)

Amplitude 
(A)

≤ 

80

±1.0% rdg.

±0.5% f.s.

±1.0% rdg.

±0.5% f.s.

±2.0% rdg.

±0.5% f.s.

80 < I 

≤ 

100

±2.5% rdg.

±0.5% f.s.

Peak 
(A peak)

|I peak| 

≤ 

110

±1.0% rdg.

±2% f.s.

±1.0% rdg.

±2% f.s.

±2.0% rdg.

±2% f.s.

110 < |I peak| 

≤ 

150

±2.5% rdg.

±2% f.s.

Phase

±1.8 deg.

DC < f < 45
(Hz):

±1.8 deg.

66 < f 

 500

(Hz):

Not defined.

CT7636 Measurement accuracy

Frequency

DC

45 

 f 

≤ 

66

(Hz)

DC < f <

 

45,

66 < f 

≤ 

1 k

(Hz)

Amplitude (A)

±2.0% rdg.

±0.5% f.s.

±2.0% rdg.

±0.5% f.s.

±3.0% rdg.

±0.5% f.s.

Peak 
(A peak)

|I peak| 

≤ 

600

±2.0% rdg.

±0.7% f.s.

±2.0% rdg.

±0.7% f.s.

±3.0% rdg.

±0.7% f.s.

600 < |I peak| 

≤ 

900

±4.0% rdg.

±0.7% f.s.

±4.0% rdg.

±0.7% f.s.

±5.0% rdg.

±0.7% f.s.

Phase

±1.8 deg.

DC < f < 45
(Hz):

±1.8 deg.

66 < f 

 1 k

(Hz):

Not defined.

CT7642 Measurement accuracy

Frequency

DC

45 

 f 

≤ 

66

(Hz)

DC < f <

 

45,

66 < f 

≤ 

1 k

(Hz)

Amplitude 
(A)

 

1800

±1.5% rdg.

±0.5% f.s.

±1.5% rdg.

±0.5% f.s.

±2.5% rdg.

±1.0% f.s.

1800 < I 

 

2000

±2.0% rdg.

±0.5% f.s.

Peak 
(A peak)

|I peak| 

 

2300

±1.5% rdg.

±1.0% f.s.

±1.5% rdg.

±1.0% f.s.

±2.5% rdg.

±1.0% f.s.

2300 < |I peak| 

≤ 

2840

±6.0% rdg.

±1.5% f.s.

±6.0% rdg.

±1.5% f.s.

±7.0% rdg.

±1.5% f.s.

Phase

±2.3 deg.

DC < f < 45
(Hz):

±2.3 deg.

66 < f 

 1 k

(Hz):

Not defined.

“Peak input” is only available in conjunction with the CM7290 

Display Unit. For more information about the combination 
accuracy with the instrument, see the instrument’s instruction 
manual.
Amplitude accuracy design value is DC < f < 5 Hz. Phase 
accuracy design value is DC < f < 10 Hz.

Zero-point Temperature 
Characteristics

Hall elements are subject to individual variation, and it is 
not possible to specify the magnitude or tendency of that 
variation. When using the sensor in an operating environment 
characterized by large temperature variations, it is 
recommended to observe variation in the zero-point under no-
input conditions. Zero-point variation affects DC offset by not 

AC mode operation. Reference examples are provided below 

to illustrate zero-point variation (23°C reference) relative 

to temperature variations for each sensor. (There is also a 

significant level of variation in characteristics among individual 

products.) The operating temperature range is −25°C to 65°C 

(−13°F to 149°F). 

(See below for example characteristics.)

CT7631

-2.0

-1.0

0.0

1.0

2.0

-40

-20

0

20

40

60

80

(A

)

(

)

No.1

No.2

No.3

Current [A]

Temperature [

°C

]

CT7636

-2.0

-1.0

0.0

1.0

2.0

-40

-20

0

20

40

60

80

(A

)

温度

(

)

No.1

No.2

No.3

Current [A]

Temperature [

°C

]

CT7642

-2.0

-1.0

0.0

1.0

2.0

-40

-20

0

20

40

60

80

(A

)

温度

(

)

No.1

No.2

No.3

Current [A]

Temperature [

°C

]

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for CT7631

Page 1: ...n and settings see the instrument s instruction manual 1 Connect the device to the instrument Instrument Align and insert the connector Align the arrow on the device s output connector with the on the compatible instrument s sensor input connector and insert the connector 2 Perform zero adjustment DC measurement Perform zero adjustment under no input conditions 3 Clamp the conductor Source Load Co...

Page 2: ...n and settings see the instrument s instruction manual 1 Connect the device to the instrument Instrument Align and insert the connector Align the arrow on the device s output connector with the on the compatible instrument s sensor input connector and insert the connector 2 Perform zero adjustment DC measurement Perform zero adjustment under no input conditions 3 Clamp the conductor Source Load Co...

Page 3: ... s Maximum cord extension length 100 m Depends on the instrument to which the device is to be connected CT7631 Measurement accuracy Frequency DC 45 f 66 Hz DC f 45 66 f 500 Hz Amplitude A I 80 1 0 rdg 0 5 f s 1 0 rdg 0 5 f s 2 0 rdg 0 5 f s 80 I 100 2 5 rdg 0 5 f s Peak A peak I peak 110 1 0 rdg 2 f s 1 0 rdg 2 f s 2 0 rdg 2 f s 110 I peak 150 2 5 rdg 2 f s Phase 1 8 deg DC f 45 Hz 1 8 deg 66 f 50...

Page 4: ... s Maximum cord extension length 100 m Depends on the instrument to which the device is to be connected CT7631 Measurement accuracy Frequency DC 45 f 66 Hz DC f 45 66 f 500 Hz Amplitude A I 80 1 0 rdg 0 5 f s 1 0 rdg 0 5 f s 2 0 rdg 0 5 f s 80 I 100 2 5 rdg 0 5 f s Peak A peak I peak 110 1 0 rdg 2 f s 1 0 rdg 2 f s 2 0 rdg 2 f s 110 I peak 150 2 5 rdg 2 f s Phase 1 8 deg DC f 45 Hz 1 8 deg 66 f 50...

Reviews: