background image

Measurement

Pre-operation Inspection

Verify that the device operates normally to ensure that no damage 

occurred during storage or shipping. If you find any damage, contact 

your authorized Hioki distributor or reseller.

Check Items

Solution

Is there any wiring break 

involving the output 

connector or sensor base?

If any, proper measurement cannot be 

made. Discontinue use and contact your 

authorized Hioki distributor or reseller.

Is the cable insulation 

torn?

If there is any damage, electric shock 

may result. Discontinue use and have 

the sensor Is there any cracking or 

repaired.

Is there any cracking or 

damage on the jaw?

 CAUTION

Do not place conductors carrying currents with frequencies 

of 10 kHz or higher in close proximity to the jaw, even if the 

device does not clamp them. Current flowing through nearby 

conductors may cause self-heating of the jaw, damaging the 

device.

This device has the protective resistance (output resistance) 

in the signal output circuit. To monitor the output signal, it is 

recommended to use a measuring instrument, such as digital 

multimeter, that has an input resistance of 1 M

Ω

 or higher.

Procedures

Attach the device to the host instrument that is

powered off.

Turn on the host instrument.

Perform a demagnetization (DEMAG) and/or zero-

adjustment (0 ADJ) (if needed).

Refer to “Demagnetization (DEMAG) and Zero-adjustment 

(0 ADJ).”

Release the jaw lock.

Open the jaw.

Clamp only one conductor to be measured and close

the jaw.

• 

• 

Check that the tips of jaw are engaged each other.

If the current direction indication is oriented against the

current direction, a polarity of output signal is reversed.

Source side

Load 

side

Conductor

Current 

direction 

indicator

Current 

direction

OK

NO

NO

Lock the jaw.

Start measurement.

On completion of the measurement,

en emove the device 

from the conductor.

str

10 

Turn off the host in ument.

11 

Detach the device from the host instrument.

Demagnetization (DEMAG) and Zero-Adjustment 

(0 ADJ)

Immediately after the power is turned on or an over-current 

exceeding the rated current is input, an offset can be output. 

Because the offset behaves as an error during DC current 

measurement, perform a demagnetization and/or zero-adjustment, 

going through the following procedure.

Open the jaw and press

the demagnetization 

(DEMAG) button.

Demagnetization (DEMAG) button

Open and close the jaw several times and check if

the offset output is stabilized, observing the host 

instrument display.

Lock the jaw.

Perform a zero-adjustment by turning 

the zero-adjust knob (0 ADJ), observing 

the offset output displayed on the host 

instrument display. 

Notch

No zero-adjustment can be performed while any current is input.

• 

• 

• 

• 

• 

Because offset output varies with the surrounding environment

and the ambient temperature (terrestrial magnetism and other

devices that generate magnetic fields), perform the zero-

adjustment at the same location at which measurements will be

made.

If the device is connected to an host instrument with a zero

correction function, align the notch of the zero-adjustment knob

with the upper center.

Mechanical shock, for example from dropping the instrument,

may cause the offset to shift.

If no zero-adjustment can be completed, perform

demagnifications (DEMAGs) several times with the jaw

remaining closed.

When measuring a DC or low-frequency (1 kHz or less) current

of small magnitude, you can increase sensitivity on a relative

basis by wrapping the conductor several times around the jaw.

Wrapping the conductor 10 times will cause the device to output

a signal equal to 10 times the measured current.

•  In the high-frequency domain,

when the high-potential side 

of a circuit (the High side) is 

clamped, measured values 

may be affected by common-

mode noise.

Source

Load

Current

High

Low

•  1.5-MHz harmonic noise arising from the operating principle can

be emitted.

•  Place the clamped conductor to be measured as close to the

jaw center as possible. Separate other conductors from the

device as far away as possible. Measuring large currents with a

frequency of 1 kHz or more or locating unclamped conductors

carrying 500 A or more of such high-frequency currents close to

the jaw can cause increases in errors or waveform distortions.

Specifications

Accuracy

f.s.

Maximum display value or scale length (This is usually the 

maximum value of the currently selected range.)

rdg.

Reading value (The value currently being measured and 

indicated on the measuring instrument)

The specs of the CT6845-05 are the same as those of the CT6845 

except the output connector and options.

Operating 

environment

Indoors, Pollution Degree II, altitude up to 

2000 m (6562 ft.)

5

6

www.calcert.com

[email protected]

1.800.544.2843

0

5

10

15

20

25

30

Reviews: