18
Manual Hold / Auto Hold
Measurement function
Fluctuation range
Threshold value
AC current
DC current
AC+DC current
20.00 A range: within 100 counts
600.0 A range: within 120 counts
2000 A range: within 40 counts
20.00 A range: 100 counts
600.0 A range: 120 counts
2000 A range: 40 counts
AC voltage
DC voltage
(excluding the 600.0 mV
range)
AC+DC voltage
6.000 V/ 60.00 V/ 600.0 V range:
within 120 counts
1000 V range: within 20 counts
1500 V range: within 30 counts
6.000 V/ 60.00 V/ 600.0 V range:
120 counts
1000 V range: 20 counts
1500 V range: within 30 counts
Resistance, Continuity
600.0
ȍ
/6.000 k
ȍ
/60.00 k
ȍ
/600.0 k
ȍ
range: within 100 counts
600.0
ȍ
/6.000 k
ȍ
/60.00 k
ȍ
/600.0 k
ȍ
range: 4900 counts
Diode
1.800 V range: within 40 counts
1.800 V range: 1460 counts
The auto hold function only operates for the above measurement functions.