42
Read Most Recent Measurement Value
Syntax
Query
:FETCh?
Response
<Total judgment result>,<Measurement value>,<Judgment result>
,<Measurement value>,<Judgment result>, …
See: “Measurement Value Formats” (p.40)
Description
Reads the most recent impedance and voltalge measurement. No trigger occurs.
See:
Data Exporting Methods (p.44), Triggering (p.37)
For (R,X,V,T) measurement, the response is as follows.
<Total judgment result> = PASS/FAIL/OFF
<Measurement value> = R measurement value
<Judgment result> = R measurement comparator judgment result <HI/IN/LO/OFF>
<Measurement value> = X measurement value
<Judgment result> = X measurement comparator judgment result < HI/IN/LO/OFF >
<Measurement value> = V measurement value
<Judgment result> = V measurement comparator judgment result < HI/IN/LO/OFF >
Example
:MEAS:VAL 1
.... Sets to return the Measured value.
:FETC?
+1.02500E-01,+1.02800E-01,+3.00
:MEAS:VAL 3
.... Sets to return the Measured value and Judgment result.
:FETC?
+1.02500E-01,IN,+1.02800E-01,IN,+3.00,IN
:MEAS:VAL 7
.... Sets to return the Measured value, Judgment result and Total judgment result.
:FETC?
PASS,+1.02500E-01,IN,+1.02800E-01,IN,+3.00
Read Temperature Measurement Value
Syntax
Query
:FETCh:TEMPerature?
Response
<Measurement value>
See: “Measurement Value Formats” (p.40)
Description
Reads the last (most recent) temperature measurement value.
Example
:FETC:TEMP?
+2.01
Summary of Contents for BT4560
Page 57: ...54 2 Enter the sample program into the code editor 3 Select Save All from the File menu Click...
Page 60: ...57...
Page 61: ...58...
Page 62: ...59...