37
(24) Triggering
Relationship between Trigger Source and Continuous Measurement Operation
Operation depends on the continuous measurement setting (
:INITIATE:CONTINUOUS
) (p.38) and the
trigger source setting (
:TRIGGER:SOURCE
) (p.38) as follows.
See:
“5 Data Exporting Methods” (p.44)
Measurement
Flow
Continuous Measurement Command-Specific Settings
:INITIATE:CONTINUOUS ON
:INITIATE:CONTINUOUS OFF
Trigger
Source
:T
RI
G
G
E
R:
S
O
URCE
I
M
M
Free-Run state. Measurement continues
automatically.
Sample Delay
Measurement
Calculation
Measured Value Output
Start measurement by
:INITIATE
(or
:READ?
)
command.
Sample Delay
Measurement
Calculation
Measured Value Output
Idle State
:INITIATE
or :READ?
:T
RI
G
G
E
R:
S
O
URCE
E
X
T
Triggered by TRIG signal or
*TRG
command.
After measurement, enters the trigger wait
state.
Sample Delay
Measurement
Calculation
Measured Value Output
Trigger Wait State
TRIG signal
or *TRG
Issue
:INITIATE
(or
:READ?
) command to
wait for trigger.
Triggered by TRIG signal.
After measurement, enters the idle state.
Sample Delay
Measurement
Calculation
Measured Value Output
Trigger Wait State
TRIG signal
Idle State
:INITIATE
or :READ?
The
:INITIATE:CONTINUOUS OFF
can only be set by Remote command.
If this has been set to OFF, when operation is returned to the Local state or power is turned off,
the
:INITIATE:CONTINUOUS ON
state occurs when power is turned back on.
See:
“Return to Local Control” (p.35, p.11)
or Exporting measured values: “Data Exporting Methods” (p.44)