It is necessary to conduct an earth voltage check before
measuring the earth resistance.
• When the instrument is connected to the Line-Earth section of
the live conductor with the MEASURE key pressed down, the
leakage breaker will be triggered and the power supply will be
cut off. Do not connect the instrument to the Line-Earth sec-
tion. Accidentally connecting to the live conductor will trigger
the protection circuit. It takes approximately 5 seconds before
measurement can be conducted again.
• As the earth leakage circuit breaker and leakage relay with
current sensitivity below 5 mA may operate, do not use this
measurement method in places where they are placed.
• The AC resistance measurement function is a function to
measure the earth resistance of a small inductance. There-
fore, accurate measurements may not be obtained when
measuring resistances with an inductance of more than 3 mH
connected in series.
1.
Set the Function selector to Earth Voltage.
2.
Connect the black earth lead to the earth electrode you
wish to measure.
2
3
1,5
4
6
Summary of Contents for Analog MO HiTESTER IR4011
Page 2: ......
Page 4: ...ii 2 6 1 D 22 2 6 2 B 26 2 7 28 2 8 30 3 31 4 41 4 1 41 4 2 41 4 3 44 4 4 L9788 10 45 1...
Page 8: ...4 IEC 61010 DC AC...
Page 9: ...5 EC EU WEEE P...
Page 10: ...6 CAT III IEC61010 CAT CAT CAT CAT CAT II CAT CAT III CAT IV...
Page 11: ...7 p 33 IH...
Page 12: ...8 AC600 V IR4030 AC150 V 1000 V 600 V IR4000 OFF...
Page 13: ...9 IP40 4 1 0 mm 1 0 mm 0 0 50 L9787 L9788 11 EN60529 IP40...
Page 14: ...10...
Page 15: ...1 11 3154 1 1 1 1 2 LED 25 V 500 V LED 35 mm LED L9788 10 LED...
Page 16: ...1 12 1 3 MEASURE IR4042 0 01 M LIGHT IR4082 L9788 10...
Page 17: ...1 13 OPEN 1 4 1 2...
Page 18: ...1 14 CAT CAT CAT OFF CAT I p 6 L9787...
Page 19: ...2 15 2 OFF V MEASURE 2 1 1 p 44 2 EARTH LINE 3 OFF 2 2 1 OFF BATT p 44 2...
Page 22: ...18 2 2 4 p 21 MEASURE OFF...
Page 23: ...2 19 2 4 1 1 2 3 5 4 1 IR4011 4015 IR4030 4042 IR4082 2 3 4 MEASURE 5 6 p 21...
Page 24: ...20 2 2 4 2...
Page 25: ...2 21 2 5 1 MEASURE OFF 2 3...
Page 26: ...22 2 IR4082 2 6 1 D 2 6 IR4082 100 Rx Ro...
Page 27: ...2 23 MEASURE 5 mA 5 mA 3 mH 2 3 1 5 4 6 1 2...
Page 28: ...24 2 IR4082 IR4082 3 4 OK 5 1 10 6 7 EARTH 8 Rx Ro...
Page 29: ...2 25 PV D PV D PV 300 V C DC AC D D B...
Page 30: ...26 2 2 6 2 B B 100 A 5 A B 2 A B A B B B 2 B 3 A A A B C D 2 3 6 4 1 5...
Page 31: ...2 27 1 2 3 4 5 1 10 6 7 8 A B...
Page 32: ...28 2 2 7 AC 600 Vrms IR4030 AC150 Vrms MEASURE...
Page 33: ...2 29 1 2 3 4 EARTH LINE MEASURE 1 V 2 3 LINE 4...
Page 34: ...30 2 OFF MEASURE 15 OFF 2 8 MEASURE OFF...
Page 45: ...4 41 4 4 1 p 42 4 2...
Page 46: ...4 42 1 73 4 3 3 MEASURE EARTH p 17 0 M p 44 p 44 MEASURE 0 M 1...
Page 47: ...4 43 2 MEASURE p 17 MEASURE 2 1 1 1 10 C 1 4 IR4000 1000 V 1000 M p 1...
Page 48: ...4 44 4 3 OFF 1 OFF 2 3 4 4...
Page 49: ...4 45 L9788 10 HIOKI L9788 90 4 4 L9788 10 1 OFF 2 7 mm 3 L9788 10 0 3 N m 4...
Page 50: ...4 46...
Page 51: ...1 1 Rx V I V V I 2 IR4000 EARTH LINE JIS C 1302 LINE EARTH...
Page 52: ...2 IR4000 1 2 500 V 600 V 10 1 1345 1 1345 2 6 JISC1302...
Page 54: ...4...
Page 55: ...IR4011 IR4012 IR4013 IR4014 IR4015 IR4030 IR4031 IR4032 IR4033 IR4041 IR4042 IR4082...
Page 56: ......
Page 58: ......
Page 86: ...Chapter 2 Measurement Procedures 26...
Page 101: ......
Page 102: ......
Page 103: ......
Page 104: ......