14.2 External I/O
386
In the sampling timing of the following modules, there is an offset from the sync
signal timing.
Input Module
Offset from sync signal
Model 8956 Analog Unit
Within -50 ns
Model 8957 High Resolution Unit
Within -500 ns
If you want to change the measurement conditions
During synchronized measurement sampling, settings such as the time base and
sampling rate cannot be changed.
If you need to change these settings, stop measurement by all of the connected
instruments before making the changes. After making the changes, select the
[SYNC Start… ]
button again in the Ext Term Settings Screen on the master
instrument.
When power is restored after a power outage during synchronized mea-
surement
Select the
[SYNC Start… ]
button in the Ext Term Settings Screen on the master
instrument.
Synchronized measurement does not restart if you do not select the button. Mea-
surement restarts if the Auto-Resume function (p. 355) is on, but no synchroniza-
tion is conducted with the connected instruments.