6.7 Triggering by Analog Signals
150
A trigger can be applied when the input signal crosses a specified trigger level
(voltage threshold). The direction in which the signal crosses the threshold is
specified by the trigger slope setting (rising
, falling
or both
).
6.7.2 Triggering When Crossing a Voltage Threshold
(Level Trigger)
Apply a trigger when the input sig-
nal exceeds 200 mV or falls below
–600 mV.
Trigger Source: OR
[Expanded] setting (Apply multiple
triggers to one channel)
No.1
Type......... Level (Level Trigger)
Ch ............ 1-1 (Unit 1 – Channel 1)
Level ....... 200 mV
Slope
(rising)
No.2
Type......... Level (Level Trigger)
Ch ............ 1-1 (Unit 1 – Channel 1)
Level ....... –600 mV
Slope........
(falling)
Level Trigger
([Normal]/[Expanded] Setting)
To open the screen: Press the
SET
key
Select
with the
SUB MENU
keys
Trigger Settings screen
See
Screen Layout (p. 35), To set from the Waveform screen (p. 167)
Operating Key
Procedure
1
CURSOR
Move the highlight cursor to a trigger number to
be set.
For the [Normal] setting, match the number of the
channel to which the trigger applies.
2
Select the trigger type.
CURSOR
Move the cursor to the
[Type]
item.
F1 to F8
Select
[Level]
(Level Trigger).
3
Select a channel (only with the [Expanded] setting).
CURSOR
Move the cursor to the
[Ch]
item.
F1 to F8
Select a channel to which the trigger applies.
4
Select the trigger level.
CURSOR
Move the cursor to the
[Level]
item.
F1 to F8
or
SCROLL
Specify the trigger threshold voltage.
5
Select the trigger slope.
CURSOR
Move the cursor to the
[Slope]
item.
F1 to F8
Select either choice.
A trigger occurs when the signal crosses the
threshold on the upslope (rising edge
).
(default setting)
A trigger occurs when the signal crosses the
threshold on the downslope (falling edge
).
A trigger occurs when the signal crosses the
threshold on either the upslope or downslope
(
).
Note: The trigger point occurs one sample af-
ter the criterion is met.
Timebase 2 sampling data is also one sample
later than that of Timebase 1.
1
200 mV
0 V
-600 mV
T
Trigger No.
Setting Example
Start
Measurement
T
If the signal is outside of the
specified threshold levels at
the start of measurement, no
triggering occurs.
Triggering occurs when the signal
crosses the threshold.
2
3
4
6
7
5
Summary of Contents for 8860-50
Page 2: ......
Page 12: ...Contents x...
Page 18: ...Reading this Manual 6...
Page 24: ...1 3 Interconnection and Block Diagrams 12...
Page 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Page 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Page 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Page 406: ...14 2 External I O 394...
Page 489: ......
Page 490: ......