Exicor MicroImager User Manual
17
2
System Operating
Requirements/Setup
Component Placement
The computer and monitor need to be placed within 0.5 meter of the system to
facilitate connection of all hardware control cables.
The system uses a fan to keep internal electrical components located in the base of
the system cool. The system must be placed such that airflow from the rear panel is
not restricted.
Power Connections
The Exicor
®
MicroImager system is designed to use an external DC power supply
that outputs 24VDC while accepting 100-240VAC, 50/60 Hz. User adjustment
should not be necessary for any internal components due to changes in power.
The Exicor
®
MicroImager system requires a single power outlet. The computer and
monitor need to be powered separately from a user-supplied wall outlet or outlet
strip.
Calibration
The Exicor
®
MicroImager has been calibrated by Hinds Instruments, Inc. prior to
shipment. The instrument should require no further adjustment for most applications.
Regularly zeroing the system will ensure the best low level sensitivity for all
measurement parameters.
Note:
All systems sold by Hinds Instruments are currently installed and
supported only through Hinds Instruments, Inc. and authorized
distributors.
Summary of Contents for EXICOR MicroImager
Page 8: ...Table of Tables viii Exicor MicroImager User Manual ...
Page 12: ...Electrical Safety 12 Exicor MicroImager User Manual ...
Page 16: ...General Dimensions 16 Exicor MicroImager User Manual ...
Page 18: ...Calibration 18 Exicor MicroImager User Manual ...
Page 23: ...Chapter 3 Operating the System Exicor MicroImager User Manual 23 ...
Page 36: ...Calibration Procedure 36 Exicor MicroImager User Manual Figure 20 ...
Page 47: ...Chapter 4 Software Interface Exicor MicroImager User Manual 47 ...
Page 49: ...Chapter 4 Software Interface Exicor MicroImager User Manual 49 ...
Page 54: ...CE Approval 54 Exicor MicroImager User Manual ...