187692(10) - EN - P280 Mattress Instructions for Use
Page 7
D
e
st
in
at
io
n
Spec
ifica
tions
Connect the device to the nearest wall plug in order to
leave the shortest possible length of cable on the floor.
In compliance with standards relating to electromagnetic
interference for medical devices, this device does not
interfere with other medical devices nor is it susceptible to interference when
combined with other medical devices that also comply with the electromagnetic
standards in place.
However, some devices, particularly older ones that do not comply with the
electromagnetic compatibility standards, may undergo interference or may
themselves interfere with the functionality of this product.
The users of such devices are responsible for ensuring that any malfunctions will
not endanger the patient or any other person.
When intravascular or intracardiac connections are in use, the electric potentials of
all the unprotected metal parts of the appliance and the bed need to be equalized.
This label indicates that
oxygen tents must never be used
and that
only the use of nasal tubes and oxygen masks is authorized. For
reasons of safety, masks and tubes should always be kept at a higher
level than the mattress support platform.
Complying with Conditions for Transport, Storage and Use
The device must be stored in its original packaging:
•
Protected against light and damp;
•
At least 10 cm above floor level to prevent fluid ingress;
•
Protected against dust;
•
Outside passageways.
Never stack more than 5 mattresses.
Symbol
Features
Use
Transport, Storage
a
a. Applicable only if the device is stored in its original packaging.
The device is designed for indoor use only.
Tempe10°C to +40°C
-30°C to +50°C
Hygrometry 30% - 85%
20% - 85%
Atmospheric
Pressure
700 hPa - 1,060 hPa
700 hPa - 1,060 hPa