![Hillrom 900 Series Instructions For Use Manual Download Page 18](http://html.mh-extra.com/html/hillrom/900-series/900-series_instructions-for-use-manual_2134308018.webp)
Page 12
171051(12) - Hillrom™ 900 Electric Bed Instructions for Use
Low position (150 diameter casters
) (h)
/439 mm
High position (double-band 125 diameter casters
) (h)
High position (125 diameter casters
) (H)
/788 mm
High position (double band 150 diameter casters
) (H)
/800 mm
High position (150 diameter casters
) (H)
/808 mm
Chassis clearance (double-band 125 diameter casters
) (C) 150 mm
Chassis clearance (125 diameter casters
) (C)
Chassis clearance (double band 150 diameter casters
) (C) 195 mm
Chassis clearance (150 diameter casters
) (C)
Head section
i
incline
+ 65°
incline*
+ 28 °
Foot section
incline*
- 3° to -22°
Trendelenburg/Reverse Trendelenburg
+ 17°/- 17°
Emergency Trendelenburg (“Shock”)
- 12 °/30 sec
Maximum patient weight for the SWL version: 220 kg
155-185 kg
j
Maximum patient weight for the SWL version: 250 kg*
185-215 kg
k
Bed weight LI900B2
l
(no mattress or accessories)
144 kg
Bed weight LI900B2 (no mattress or accessories)
120 kg
Bed weight LI900B3 (no mattress or accessories)
170 kg
Maximum temperature of applied parts at 40° C
56,5° C
Unweighted peak acoustic pressure levels
<120 dB
Maximum measured level of weighted acoustic pressure
42 dBA
a. These are average values, which may vary according to manufacturing tolerances..
b. Bed fitted with AD271B siderails
c. Bed fitted with AD272B siderails
d. Dimensions in mm.
e. An antistatic version is also available.
f. LI900B3 model
g. Incompatible with the CFS 250 kg version, LI900B2 with half-siderail.
h. LI900B2 model
i. Maximum incline in relation to the sleep surface
j. SWL 220 kg / the maximum patient weight varies according to the mattress and accessories
used
- 155 kg as per EN60601-2-52 (acute care or intensive care)
- 185 kg as per EN60601-2-52 (other environments).
k. SWL 250 kg / the maximum patient weight varies according to the mattress and accessories
used
- 185 kg as per EN60601-2-52 (acute care)
- 215 kg as per EN60601-2-52 (other environments).
l. LI900B2 models with half-siderails
Features
Value