4990824_030_02 – 2071995 – 12-DEC-2019
61
Use
–
If the patient weighs more than 360 kg / 793 lbs, the operating
tabletop must not be tilted by more than 15°. The operating
tabletop stops automatically at 15° and must not be tilted
beyond this.
6.10.6
Longitudinal slide
NOTICE
Danger of collisions when moving the operating tabletop
Accessories on the side rail may collide with the operating table
column during longitudinal travel.
•
Monitor the operating table's movement and stop the function
before any potential collision.
Longitudinal slide, head end:
Press the [i16] key on the remote control or column keypad.
Longitudinal slide, foot end
Press the [i17] key on the remote control or column keypad.
This function is restricted in the following situations:
–
The longitudinal travel is locked when the extension adapter is
attached to the operating tabletop.
–
If the leg section or back section is tilted downward,
longitudinal travel may be restricted. The adjustment range
depends how far the leg section or back section is tilted
downward. If a larger adjustment range is required, the leg
section or back section must be moved upward.
–
The longitudinal slide towards the foot end of the operating
table is restricted to 100 mm if one of the following tabletop
sections is attached at the operating tabletop’s foot-end hook
coupling:
You will hear an error message if longitudinal travel is in the
impermissible range. In this case, longitudinal travel is only
possible toward the head end until the permissible range is
reached again.
53
54
63
Item
Product designation
Part
number
Table top segment Carbon1200 H V
Summary of Contents for 4080300
Page 2: ...This page is intentionally blank ...
Page 4: ...Mobile operating table 4 4990824_030_02 2071995 12 DEC 2019 This page is intentionally blank ...
Page 6: ...Mobile operating table 6 4990824_030_02 2071995 12 DEC 2019 This page is intentionally blank ...
Page 10: ...Contents 10 4990824_030_02 2071995 12 DEC 2019 This page is intentionally blank ...