2.4 Manual Hilow Function Does Not Raise or Lower Correctly
Chapter 2: Troubleshooting Procedures
Resident
TM
LTC Bed Service Manual (man136rc)
Page 2 - 23
2
2.4
Manual Hilow Function Does Not Raise or Lower Correctly
1. The bolt and nut securing the hilow crank handle to the hilow drive
assembly are present and secure.
Yes
No
↓
→
Ensure the bolt (A) and nut (B) securing the hilow crank handle
to the hilow drive assembly are present and secure (see figure 2-
14 on page 2-23). If the bolt or nut is missing, consult chapter 5
“Parts List” for the part number(s) of the appropriate
replacement part(s).
Figure 2-14. Crank Handle Assembly
2. The hilow support bracket (A) is present and secure (see figure 2-15 on
page 2-23). The pin (B) and rue ring cotters (C) securing the hilow drive
assembly to the bed frame assembly are also present and securely fastened.
Yes
No
↓
→
Figure 2-15. Connection of the Manual Hilow Drive Assembly to the Main Bed Frame
B
A
m136_112
A
136 113
B
C
m136_113