2.2 Siderail Mechanical Malfunction
Chapter 2: Troubleshooting Procedures
Page 2 - 20
Resident
TM
LTC Bed Service Manual (man136rc)
2
2.2
Siderail Mechanical Malfunction
1. The siderail moves smoothly from/to the stored position.
Yes
No
↓
→
Check that the slide rods are not bent, guide pins (A) enter the
aligned holes through the channel assembly, and engage the
holes in strap (B) (see figure 2-13 on page 2-20). Repair or
replace the damaged parts (see “Siderails” on page 4-37).
This solves the problem.
Yes
No
↓
→
Go to step 5.
Figure 2-13. Siderail Assemblies (Shown with Electric Siderail Controls Installed)
2. The siderail rotates smoothly to the up position and latches correctly.
Yes
No
↓
→
Go to step 5.
3. In the latched position, the siderail does not move inward where the link
arms can contact the bed frame.
Yes
No
↓
→
Go to step 7.
A
B
C
D
m136_017
Summary of Contents for P870
Page 13: ...Table of Contents Page xii ResidentTM LTC Bed Service Manual man136rc NOTES...
Page 99: ...Chapter 3 Theory of Operation Page 3 2 ResidentTM LTC Bed Service Manual man136rc 3 NOTES...
Page 163: ...Chapter 5 Parts List Page 5 2 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...
Page 165: ...Warranty Chapter 5 Parts List Page 5 4 ResidentTM LTC Bed Service Manual man136rc 5 NOTES...