2.27 OB/GYN Stretcher—Foot Section Does Not Release Correctly
Chapter 2: Troubleshooting Procedures
2.27 OB/GYN Stretcher—Foot Section Does Not Release Correctly
1.
Another part of the stretcher interferes with the foot section assembly.
Yes
No
↓
→
2.
Remove the interference.
This solves the problem.
Yes
No
↓
→
3.
Go to “Final Actions” on page 2-11.
4.
The stretcher is in the standard position (non-examination).
Yes
No
↓
→
Move the patient platform into the standard position. If this
solves the problem, go to “Final Actions” on page 2-11.
Otherwise, continue to step 5.
5.
Make sure the release handle and cable function operate correctly. All
mechanisms work correctly.
Yes
No
↓
→
Adjust the cable assembly for the release handle. If this solves
the problem, go to “Final Actions” on page 2-11. Otherwise,
continue to step 6.
6.
Replace the foot section assembly. The foot section operates correctly.
Yes
No
↓
→
Call Technical Support.
7.
Go to “Final Actions” on page 2-11.
Page 2 - 40
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Service Manual (144386 REV 2)